option can prevent trace levels of acid in the gas from
compromising the sensor. The maximum allowable levels of acid component in
the sample stream are the following:
Measuring Range
CO
2
*
SO
2
H
2
S
NO
X
Cl
2
HCl
0-100 ppm
0.2
200
200
200
100
100
0-500 ppm
0.2
200
200
200
100
100
0-1000 ppm
0.4
500
500
500
200
200
0-5000 ppm
0.6
1000
1000
1000
400
400
0-10,000 ppm
0.8
1500
1500
1500
800
800
0-5%
2.0
2600
2600
2600
1400
1400
0-10%
4.0
4000
4000
4000
2000
2000
0-25%
6.0
6000
6000
6000
3000
3000
*Concentrations of CO
2
are in percent. One percent is equivalent to 10,000 ppm.
Table 2: Maximum Acid Gas Concentrations with Stab-El
(in ppm)
Contact Servomex for recommendations on using the Stab-El
sensor on acid
gases other than those listed above.
If the acid gas components in your gas exceed the limits in Table 2 several
approaches can be taken:
An acid scrubber can be placed up stream of the analyzer to remove the
bulk of the acids, and the Stab-El
option will allow the analyzer to cope
with the remaining trace levels.
The analyzer duty cycle can be adjusted to limit its exposure to acids and to
allow the Stab-El
option to “keep up”.
A dilution system can be added to limit acid concentration.
Consult Servomex for further recommendations.
Summary of Contents for DF-130E
Page 1: ...Oxygen Analyzer DF 150E OPERATOR MANUAL...
Page 10: ......
Page 13: ...Specifications DF 150E 9 Figure 1 DF Series Oxygen Analyzer...
Page 14: ......
Page 44: ......
Page 56: ......
Page 84: ......
Page 87: ...Index DF 150E 83 U Upsets Oxygen Shock 44 V Voltage 15 46 checking 15 W Warranty 79...