8 DF-150E
Specifications
Response Time
Responds instantaneously to O
2
change. Equilibrium time depends on specific
conditions. Typically less than 10 seconds to read 90% of a step change.
Accuracy
At constant conditions, the greater of +/- 3% of reading or 0.5% of
measurement range, but not lower than 0.05% of full analyzer range for three
range analyzers.
Sample Gas Compatibility
Standard Sensor
All inert and passive gases including N
2
, H
2
, CO, freons, light hydrocarbons,
etc.
Sensor with Stab-El
Neutralizes contaminants including acids such as CO
2
, H
2
S, Cl
2
, NO
X
, SO
X
,
etc. (See page 19 for concentration limits).
Sample Requirements
Gas phase, non-condensing, sample temperature: -17 to 66°C (0 to 150°F).
Flow rate 1.0 to 2.0 scfh at 0.2 to 1.0 psig pressure.
Ambient Temperature Requirements
Room/cabinet temperature: 0 to 50°C (32 to 120°F)
Physical Dimensions
Overall Dimensions:
20.3cm w X 21.6cm h X 27.9cm d
Weight:
6 kg
Overall Dimensions:
8.0”w X 8.5”h X 11”d
Weight:
15 lbs
Miscellaneous
Store and operate between 0°C and 45°C (32°F and 112°F).
100-120/200-240 VAC, 50-60Hz, 1.0/0.5A maximum.
Nominal power consumption: 20 Watts
ETLc Approved
Summary of Contents for DF-130E
Page 1: ...Oxygen Analyzer DF 150E OPERATOR MANUAL...
Page 10: ......
Page 13: ...Specifications DF 150E 9 Figure 1 DF Series Oxygen Analyzer...
Page 14: ......
Page 44: ......
Page 56: ......
Page 84: ......
Page 87: ...Index DF 150E 83 U Upsets Oxygen Shock 44 V Voltage 15 46 checking 15 W Warranty 79...