Shipping the Analyzer DF-150E
59
13 Shipping the Analyzer
If it becomes necessary to ship the analyzer for service contact your local
Servomex Business Center to obtain a Return Material Authorization (RMA)
number.
Be sure to mark the Return Material Authorization number on the outside of
the shipping carton and on the packing list.
CAUTION
Do not ship the analyzer with liquid in the sensor. Serious damage
can result and the warranty will be voided.
NOTE
The analyzer should be carefully packed in its original packing
materials if possible.
13.1Draining the Sensor
Remove the Sensor
1. Open the front door.
2. Turn power off.
3. Disconnect the electrical cable that runs to the sensor.
4. Loosen the two, knurled thumb screws in front of the sensor.
5. Remove the bulkhead nut from the inlet gas connection on the rear of the
analyzer.
6. If the analyzer is equipped with the stainless steel outlet tubing option,
remove the nut from the outlet bulkhead fitting.
7. Carefully slide the sensor out of the analyzer a few inches.
8. Remove the quick disconnect at the top of the flowmeter. (Skip this step if
the analyzer is equipped with the stainless steel outlet tubing option.)
9. Finish removing the sensor from the cabinet.
10. Carefully un-screw the lid from the sensor.
11. Tip the sensor, carefully capture the liquid, and dispose of it appropriately.
12. Rinse sensor with distilled or de-ionized water.
Summary of Contents for DF-130E
Page 1: ...Oxygen Analyzer DF 150E OPERATOR MANUAL...
Page 10: ......
Page 13: ...Specifications DF 150E 9 Figure 1 DF Series Oxygen Analyzer...
Page 14: ......
Page 44: ......
Page 56: ......
Page 84: ......
Page 87: ...Index DF 150E 83 U Upsets Oxygen Shock 44 V Voltage 15 46 checking 15 W Warranty 79...