Soluciones y Tecnologías de Control Embebido S.A.P.I. de C.V.
www.sepradel.com [email protected] +52 1 833-389-4054
24
you must also set the ideal delay between the sync pulse of the flash lamp and the pulse destined
for the spectrometer. As a general rule, it is recommended to first adjust the delay of channel “C”
connected to the Q-Sw, until obtaining laser plasma or the desired laser power and then program
channel “B” with the same delay found for the Q-Switch and from said value increase or decrease
it until optimizing the obtaining of the optical spectrum.
In the following three figures, one of the three possible laser outputs is used as synchronism
(some lasers only provide the “LAMP” signal), since the LAMP and Q-Switch control is carried out
with the delays characteristic of the laser console.
Fig. 24. Synchronism with flash lamp, lamp and Q-Switch controlled by the laser itself
Fig. 25. Synchronism with Q-Switch, flash lamp and Q-Switch Controlled by the laser itself