background image


 

Cleaning 

Proper maintenance of a probe is important. If the probe is not cleaned properly and regularly, probe 
malfunction can occur due to a polluted diaphragm or ISFET. The cause for this pollution is usually the 
sample. 

Pollution on/or blockage of the sensor and reference electrode diaphragm surface is the most likely 
cause for probe failure. If any of the following events occur: low slope, drift, instability of the reading, 
slow calibration, probe will not calibrate, pH value doesn’t change as expected when changing samples, 
start the cleaning procedure as described below and perform a revitalization. 

Probe wear is another cause for probe failure. Probe wear is often, but not necessarily, preceded by a 
period of declining calibration slope values. Probe wear is dependent on how the probe is used, stored. 
Worn probes need to be replaced.  

 

3.1  Daily cleaning 

The appropriate cleaning frequency is dependent on the type of sample being measured.  A good rule 
of thumb to use when sampling colored liquids is when the reference diaphragm is no longer white, 
the probe should be cleaned.  

Both the surface of the chip (the shiny dot at the probe’s tip) and the diaphragm (the white surface on 
the probe tip) are to be cleaned with a brush. Place the probe in warm tap water (around 60°C / 140°F) 
with a mild detergent for 5 minutes, stir periodically. Scrub the probe tip with a soft brush and water 
with a mild detergent. After scrubbing rinse with deionized water.  

Never brush the probe tip, especially the ISFET chip, before rinsing and flushing thoroughly with water 
(tap water, demi-water etc).  Before rinsing debris and particles  may  be  on  the  sensor surface and 
brushing them into the senor may damage it. When in doubt, soak the probe for a while in warm water 
with a mild detergent. 

 

3.2  Revitalizing 

Revitalization is performed to regenerate the diaphragm in the pH probe. The diaphragm is part of the 
so called reference electrode and necessary for pH measurements. 

For best results, clean the probe first as described in section  3.1 before revitalizing. Make sure the 
probe  is  still  warm  (around  60°C  /  140°F)  and  place  the  probe  directly  (without  flushing  it  with 
deionized water or cooling it down) in a saturated KCl-solution at room temperature and let it stand 
for 20 minutes. This ‘cold’ KCl-dip will regenerate the reference system and the diaphragm.  

 

To prepare saturated KCl (potassium chloride) solution: Add KCl-granules to distilled water until 
no more KCl will dissolve. Adding 38 grams of KCl to 100 ml water is sufficient. Let this stand for at 
least two hours and decant the clear solution. Now you have saturated KCl. 

 

Summary of Contents for ConeFET

Page 1: ...SI line pH PROBE MANUAL For products ConeFET 3280 010 CupFET 3200 010 LanceFET 2270 010 LanceFET H 2274 010 MicroFET 9270 010 MiniFET 9202 010...

Page 2: ...he probe as this may cause damage and will void the guarantee All information provided in this manual is applicable at the moment of publication Sentron holds the right to change components functions...

Page 3: ...of compliance 4 1 2 Guarantee 4 1 3 Specifications 5 2 Use 6 2 1 Installation 6 2 2 Preparing for use 7 2 3 Operating tips 8 3 Cleaning 9 3 1 Daily cleaning 9 3 2 Revitalizing 9 3 3 Cleaning tips 10...

Page 4: ...n will according to its own judgment either replace or repair a defect probe within 6 months of shipment provided the cause is a defect in material of manufacturing and the defect appeared during corr...

Page 5: ...5 1 3 Specifications...

Page 6: ...obe failure or irreversible damage to the probe Samples must be aqueous solutions or semi solids and compatible with the probe s materials If information is required regarding the chemical resistance...

Page 7: ...page from a new probe is normal and will not affect the lifetime or performance of the probe Subsequently rinse with deionized or distilled water Soak the probe in pH 7 00 buffer or pH 4 00 buffer wit...

Page 8: ...late based are less susceptible to carbon dioxide contamination than pH 10 00 buffers borax or carbonate based When slope errors occur it usually indicates a failing probe or a contaminated buffer If...

Page 9: ...e to be cleaned with a brush Place the probe in warm tap water around 60 C 140 F with a mild detergent for 5 minutes stir periodically Scrub the probe tip with a soft brush and water with a mild deter...

Page 10: ...r Cleaning agents are available at Sentron ask your local dealer Do not use hydrofluoric acid acetone MEK or similar agents to clean the probe 4 Storage For short time periods 2 Days probes can be sto...

Reviews: