background image

 

2.3  Operating tips 

 

Clean your probe regularly during use. For cleaning instructions see section 3. 

 

Ensure samples or buffers are well mixed to obtain correct measurement values. This may be 
done by a magnetic stirrer or by stirring with the probe for at least 5 seconds. Stop stirring and 
record results when the read-out is stable. 

 

Make sure that the medium which is being measured, is providing a “bridge” between the ISFET 
and the diaphragm. Without a proper fluid (or semi-solid) connection between these both, no 
(stable) measurement can be performed. A good rule of thumb is to have at least the first 10 
mm from the probe tip to be immersed during measurements and calibrations. 

 

When  using  CupFET  or  MicroFET  probes,  especially  when  performing  single  droplet 
measurements, make  sure  the  medium is in contact  with both the  ISFET and the  diaphragm 
simultaneously.  The  CupFET  should  contain  the  medium  within  the  “cup”  shaped  tip  of  the 
probe. 

 

When using ConeFET or LanceFET probes with semi-solids, insert probe to desired depth, then 
rotate left and right several times and tilt to ensure sample contact. 

 

Buffer-handling: pH 7.00 buffers (phosphate-based) and pH 4.00 buffers (biphtalate-based) are 
less  susceptible  to  carbon  dioxide  contamination  than pH 10.00 buffers (borax or  carbonate 
based). When slope errors occur, it usually indicates a failing probe or a contaminated buffer. If 
slope errors occur when using a pH 10.00 buffer, try calibrating with pH 7.00 and pH 4.00 buffer. 
If a good slope is achieved, try a new bottle of pH 10.00 buffer. Buffers in a convenient twin-
neck bottle are available from Sentron or our dealers.   

 

When testing in direct sunlight or on a bright reflecting surface, please use brown, opaque or 
shielded sample containers. Very bright light might influence the performance of the sensor. 

 

Tris buffers and samples containing proteins form impermeable layers on surfaces, and require 
special attention when being used. These types of samples should be measured quickly and the 
probe  should be  rinsed thoroughly with deionized water between samples. Avoid prolonged 
immersion in samples containing Tris or proteins. When testing is complete, first clean the probe 
with water and a laboratory detergent and subsequently rinse with deionized or distilled water.  

 

Avoid prolonged immersion in samples expected to have pH-values at the ends of the specified 
pH-range. When prolonged exposure to very high or very low pH values does take place rinse 
the probe  with neutralizing agents and distilled water when the measurement is completed. 

 

 

 

Summary of Contents for ConeFET

Page 1: ...SI line pH PROBE MANUAL For products ConeFET 3280 010 CupFET 3200 010 LanceFET 2270 010 LanceFET H 2274 010 MicroFET 9270 010 MiniFET 9202 010...

Page 2: ...he probe as this may cause damage and will void the guarantee All information provided in this manual is applicable at the moment of publication Sentron holds the right to change components functions...

Page 3: ...of compliance 4 1 2 Guarantee 4 1 3 Specifications 5 2 Use 6 2 1 Installation 6 2 2 Preparing for use 7 2 3 Operating tips 8 3 Cleaning 9 3 1 Daily cleaning 9 3 2 Revitalizing 9 3 3 Cleaning tips 10...

Page 4: ...n will according to its own judgment either replace or repair a defect probe within 6 months of shipment provided the cause is a defect in material of manufacturing and the defect appeared during corr...

Page 5: ...5 1 3 Specifications...

Page 6: ...obe failure or irreversible damage to the probe Samples must be aqueous solutions or semi solids and compatible with the probe s materials If information is required regarding the chemical resistance...

Page 7: ...page from a new probe is normal and will not affect the lifetime or performance of the probe Subsequently rinse with deionized or distilled water Soak the probe in pH 7 00 buffer or pH 4 00 buffer wit...

Page 8: ...late based are less susceptible to carbon dioxide contamination than pH 10 00 buffers borax or carbonate based When slope errors occur it usually indicates a failing probe or a contaminated buffer If...

Page 9: ...e to be cleaned with a brush Place the probe in warm tap water around 60 C 140 F with a mild detergent for 5 minutes stir periodically Scrub the probe tip with a soft brush and water with a mild deter...

Page 10: ...r Cleaning agents are available at Sentron ask your local dealer Do not use hydrofluoric acid acetone MEK or similar agents to clean the probe 4 Storage For short time periods 2 Days probes can be sto...

Reviews: