Use 

The pH-sensitive element in a SI line Sentron pH probe is an ISFET semi-conductor sensor. This sensing 
element needs to be driven by an electronic circuit that is built into the Sentron SI line pH meter. This 
probe  can  only  be  used  in  combination  with  one  of  the  Sentron  SI  line  pH-meters.  Any  other 
combination will cause erroneous measurement values and irreversible damage to both probe and 
meter. 

Sentron probes are robust and build to last.  The pH probe nonetheless requires periodic maintenance. 
Please read the Operating Tips later in the this manual to ensure your probe is properly maintained. 

Do not use the probe outside the specified temperature range as this might result in probe failure or 
irreversible damage to the probe. 

Samples must be aqueous solutions or semi-solids and compatible with the probe’s materials.  

If information is required regarding the chemical resistance of the probe, please contact your local 
dealer or Sentron. 

2.1  Installation 

Push the probe’s connector into the socket on the upper side of the meter. The (white) notch on the 
probe connector needs to be aligned with the meter, having the notch facing towards the front side of 
the meter. The connector is pushed down until it clicks into the meter’s socket.  

  

The probe can be disconnected from the meter by simply pulling the probe connector. 

 

 

Connect the probe to the meter   

 

Note:  

Do not turn the probe connector once inserted into the meter’s socket. Also, make sure 
both  meter  and  probe  connector  are  clean  and  free  from  debris  before  connection. 
Do not pull the probe’s cable, always pull the connector. 

Summary of Contents for ConeFET

Page 1: ...SI line pH PROBE MANUAL For products ConeFET 3280 010 CupFET 3200 010 LanceFET 2270 010 LanceFET H 2274 010 MicroFET 9270 010 MiniFET 9202 010...

Page 2: ...he probe as this may cause damage and will void the guarantee All information provided in this manual is applicable at the moment of publication Sentron holds the right to change components functions...

Page 3: ...of compliance 4 1 2 Guarantee 4 1 3 Specifications 5 2 Use 6 2 1 Installation 6 2 2 Preparing for use 7 2 3 Operating tips 8 3 Cleaning 9 3 1 Daily cleaning 9 3 2 Revitalizing 9 3 3 Cleaning tips 10...

Page 4: ...n will according to its own judgment either replace or repair a defect probe within 6 months of shipment provided the cause is a defect in material of manufacturing and the defect appeared during corr...

Page 5: ...5 1 3 Specifications...

Page 6: ...obe failure or irreversible damage to the probe Samples must be aqueous solutions or semi solids and compatible with the probe s materials If information is required regarding the chemical resistance...

Page 7: ...page from a new probe is normal and will not affect the lifetime or performance of the probe Subsequently rinse with deionized or distilled water Soak the probe in pH 7 00 buffer or pH 4 00 buffer wit...

Page 8: ...late based are less susceptible to carbon dioxide contamination than pH 10 00 buffers borax or carbonate based When slope errors occur it usually indicates a failing probe or a contaminated buffer If...

Page 9: ...e to be cleaned with a brush Place the probe in warm tap water around 60 C 140 F with a mild detergent for 5 minutes stir periodically Scrub the probe tip with a soft brush and water with a mild deter...

Page 10: ...r Cleaning agents are available at Sentron ask your local dealer Do not use hydrofluoric acid acetone MEK or similar agents to clean the probe 4 Storage For short time periods 2 Days probes can be sto...

Reviews: