5.2 Examples
Below some communication examples on the various protocol functions are presented to illustrate
communication with the AD Converter and USB Interface modules.
5.2.1 Performing a single point calibration
pH7 calibration sequence
:
Rinse the probe with deionized water.
Place ISFET sensor and reference in the calibration pH7 buffer solution. Initiate the
calibration process by sending the S
tart Calibration
command:
CLR!<CR>
Wait for the AD Converter or USB interface module to acknowledge, receive:
082 013 013
The AD converter or USB interface module is now ready to receive the calibration pH 7
command…
Initiate the pH7 calibration by sending the calibration pH 7 command:
113!<CR>
Allow some time for the module to stabilize (this may take up to 2 minutes maximum)…
Wait for the module to stabilize, receive bytes:
003 013 010
End the calibration process, send:
QIT!<CR>
Wait for the module to confirm calibration end, receive:
084 013 010
Rinse the ISFET sensor and reference electrode with de-mineralized water…
The calibration is completed.
For performing a pH 2, 4, 10 or 12 calibration, repeat the sequence and use the appropriate
calibration command bytes. See chapter 5.1 for the command bytes of each pH buffer solution.