background image

 

 

 

 

 

 

Description 

 
Sentron  offers  a  glass-free  modular  pH-kit  designed  specifically  for  development  and  testing 
purposes. With this pH-kit, development engineers and researchers have a large degree of flexibility 
in how they integrate our proprietary ISFET pH sensor into their applications or experimental set-ups. 
Due to the modular design, the functionality can be  expanded as required and components can be 
replaced individually. 
 
The pH-kit consist of a ISFET pH sensor, Reference and Analog front-end module. These modules are 
always needed as a basis for pH measurement. Thanks to the small size of the  ISFET pH sensor and 
Reference  electrode  it  can  be  used  to  measure  small  volumes  or  develop  applications  with  small 
form  factors.  The  Analog  Front-end  module  has  an  uncalibrated  analog  pH  output  signal  with  a 
voltage output 0 — 3.3 V of ~ 52 mV / pH and pH 7 between 500 mV and 1800 mV. The PT1000 RDT 
temperature sensor in the ISFET pH sensor is wired directly to the Analog front-end module output. 
The  reference  electrode  also  connects  to  the  Analog  front-end  module  and  either  the  standard 
reference  electrode  with  porous  PTFE  diaphragm  can  be  used  or  a  suitable  custom  reference 
electrode can be attached. 
 
The AD Converter module can be attached to the  Analog front-end module. This extension module 
with microprocessor, AD Converter and galvanic isolation makes it well suited for use in embedded 
applications. The communication with the AD Converter is based on a serial RS232 interface with a 
TTL  level.    Using  a  standard  serial interface  it  is  possible  to  perform  calibrations  and  read  pH  and 
temperature values. Application of a temperature correction algorithm to the pH signal is performed 
directly by the microcontroller of the AD Converter module. The galvanic isolation provides an extra 
safety barrier and prevents ground loops.  
 
With the USB Interface extension module, which connects to the AD Converter module, it is possible 
to  request  measured  pH  and  temperature  values  from  a  laptop  or  PC  with  a  USB    port.  The  USB 
Interface extension module, with appropriate user developed software, allows the pH-kit to be used 
for applications that require direct connection to a PC, such as real time monitoring of pH values in 
an experiment or process. 
 
 

 

Summary of Contents for A120-001

Page 1: ...nverter module 4 3 3 USB Interface module 5 Calibration 6 Data communication 7 5 1 Protocol 7 5 2 Examples 8 5 2 1 Performing a single point calibration 8 5 2 2 Performing a multi point calibration 9 5 2 3 Retrieving the pH signal 10 5 2 4 Retrieving the temperature signal 10 5 2 5 Retrieving the slope values 11 5 3 ASCII table 12 E7500347 02 ...

Page 2: ... connects to the Analog front end module and either the standard reference electrode with porous PTFE diaphragm can be used or a suitable custom reference electrode can be attached The AD Converter module can be attached to the Analog front end module This extension module with microprocessor AD Converter and galvanic isolation makes it well suited for use in embedded applications The communicatio...

Page 3: ...3 Specifications ...

Page 4: ... the Analog front end module They are described below 7 Pin 1 pH signal out 52 mV pH pH 7 between 500 and 1800 mV Pin 2 3V3 Power 3V3 DC power power input 100mV Pin 3 AGND Analog Ground Pin 4 PT1000 Directly wired to the PT1000 Pin 5 PT1000 Directly wired to the PT1000 Pin 6 N C Not used Pin 7 Reference External reference electrode connection 3 2 AD Converter module The 6 pin connector of the AD C...

Page 5: ...ognize the USB interface as a USB to COM port converter Windows will install the driver automatically or download the driver from the internet In some cases the driver needs to be installed manually The latest drivers can be downloaded from the FTDI website http www ftdichip com Drivers VCP htm When Windows has installed the driver properly it will assign a COM port number to the USB Interface Thr...

Page 6: ...als The frequency of recalibration is dependent on the conditions the sensor is subjected to and it is recommended to test the calibration at the end of a measurement run to confirm the calibration is still accurate To perform a calibration place the ISFET sensor in the right buffer solution and communicate according the protocol of chapter 5 1 ...

Page 7: ...nd ASCII command bytes Decode calculation Multiply or add decimal byte position value Units A B C D1 E1 F1 G1 H1 I1 J K A 4096 B 64 C 0 001 pH byte byte byte 000 000 000 000 000 000 013 010 A B C 1 D 1 E 1 F G A 64 B 0 1 F byte byte 000 000 255 013 010 Start Calibration CLR CR 067 076 082 033 013 082 013 013 Calibration pH 2 111 CR 5 001 001 001 033 013 001 013 010 2 Calibration pH 4 112 CR 5 001 ...

Page 8: ...2 013 013 The AD converter or USB interface module is now ready to receive the calibration pH 7 command Initiate the pH7 calibration by sending the calibration pH 7 command 113 CR Allow some time for the module to stabilize this may take up to 2 minutes maximum Wait for the module to stabilize receive bytes 003 013 010 End the calibration process send QIT CR Wait for the module to confirm calibrat...

Page 9: ...he module to stabilize receive bytes 002 013 010 Rinse the probe with deionized water Place ISFET sensor and reference in the next calibration buffer solution pH 7 Initiate the pH7 calibration by sending the calibration pH 7 command bytes 113 CR Allow some time for the module to stabilize this may take up to 2 minutes maximum Wait for the module to stabilize receive bytes 003 013 010 Rinse the pro...

Page 10: ...ition A B C D E F G H I J K Received bytes 001 023 027 000 00 000 000 000 000 013 010 Protocol A 001 B 023 C 027 DEFGHIJK n a A 4096 B 64 C 1 4096 23 64 27 5595 pH value 5 595 5 2 4 Retrieving the temperature signal Retrieve temperature value Send the command bytes 777 CR Wait for the module to return the pH value receive bytes ABCDEFG Data ABCDEFG marks the byte position Values in bytes are neede...

Page 11: ... replace the ISFET sensor and or reference electrode Retrieve slope value Send the command bytes 000 CR Wait for the module to return the slope values receive bytes ABCDEFGHIJKLMN Data ABCDEFGHIJKLMN marks the byte position Values in bytes are needed for decoding Byte position A B C D E F G H I J K L M N Byte value 001 byte byte 002 byte byte 003 byte byte 004 byte byte 013 010 Decode received seq...

Page 12: ...12 5 3 ASCII table ...

Reviews: