background image

 

 

 

 

 

 

Electrical connections 

The  diagram  below  illustrates  how  the  components  of  the  pH-kit  are  connected  together.  The 
modules will only function when connected in this order. 

 

1.  

ISFET  

Reference electrode 

2.  

Analog front-end 

3.  

AD Converter 

(Optional) 

4.  

USB Interface 

(Optional) 

 
 
 

 

Modules 

and 

are always needed  to conduct pH measurements with the test kit. Modules 

3

 and 

4

 are optional. 

3.1  Analog front-end module 

The ISFET sensor is connected to the Analog Front-End module through an FFC connector. There are 
also 7 pin connections on the module. One pin (Pin 7) is for connection of the reference electrode. 
The remaining 6 serve as output pins for the Analog front-end module. They are described below:  

                                                               7

.

 

 

Pin 1.

  pH signal out 

~52 mV/pH. pH 7 between 500 and 1800 mV 

Pin 2.

  +3V3 Power 

+3V3 DC power power input +/- 100mV  

Pin 3.

  AGND

   

Analog

 

Ground 

Pin 4.

  PT1000    

Directly wired to the PT1000 

Pin 5.

  PT1000    

Directly wired to the PT1000 

Pin 6.

  N.C. 

 

Not used 

Pin 7.

  Reference

 

External reference electrode connection 

3.2  AD Converter module 

The  6 pin connector of the  AD Converter module  can be  attached directly to the  Analog Front-end 
module. The AD Converter module also had 4 output pins, which are described below:   




Pin 1.

  +5V  

 

+5V DC power input +/- 100mV 

Pin 2.

  RxD

 

 

Data input, TTL 5V level, 115k2, 8N1 

Pin 3.

  TxD

 

 

Data output, TTL 5V level, 115k2, 8N1 

Pin 4.

  GND 

 

Digital GND. 

 
The AD Converter module contains 1.5kV galvanic insulation between inputs and outputs.  

1.

2.

 

3.

4.

1.

2.

 

3.

4.

5.

  

6. 

  

 

ISFET 

sensor 

connector 

 

Analog 

front-end 

connector 

Summary of Contents for A120-001

Page 1: ...nverter module 4 3 3 USB Interface module 5 Calibration 6 Data communication 7 5 1 Protocol 7 5 2 Examples 8 5 2 1 Performing a single point calibration 8 5 2 2 Performing a multi point calibration 9 5 2 3 Retrieving the pH signal 10 5 2 4 Retrieving the temperature signal 10 5 2 5 Retrieving the slope values 11 5 3 ASCII table 12 E7500347 02 ...

Page 2: ... connects to the Analog front end module and either the standard reference electrode with porous PTFE diaphragm can be used or a suitable custom reference electrode can be attached The AD Converter module can be attached to the Analog front end module This extension module with microprocessor AD Converter and galvanic isolation makes it well suited for use in embedded applications The communicatio...

Page 3: ...3 Specifications ...

Page 4: ... the Analog front end module They are described below 7 Pin 1 pH signal out 52 mV pH pH 7 between 500 and 1800 mV Pin 2 3V3 Power 3V3 DC power power input 100mV Pin 3 AGND Analog Ground Pin 4 PT1000 Directly wired to the PT1000 Pin 5 PT1000 Directly wired to the PT1000 Pin 6 N C Not used Pin 7 Reference External reference electrode connection 3 2 AD Converter module The 6 pin connector of the AD C...

Page 5: ...ognize the USB interface as a USB to COM port converter Windows will install the driver automatically or download the driver from the internet In some cases the driver needs to be installed manually The latest drivers can be downloaded from the FTDI website http www ftdichip com Drivers VCP htm When Windows has installed the driver properly it will assign a COM port number to the USB Interface Thr...

Page 6: ...als The frequency of recalibration is dependent on the conditions the sensor is subjected to and it is recommended to test the calibration at the end of a measurement run to confirm the calibration is still accurate To perform a calibration place the ISFET sensor in the right buffer solution and communicate according the protocol of chapter 5 1 ...

Page 7: ...nd ASCII command bytes Decode calculation Multiply or add decimal byte position value Units A B C D1 E1 F1 G1 H1 I1 J K A 4096 B 64 C 0 001 pH byte byte byte 000 000 000 000 000 000 013 010 A B C 1 D 1 E 1 F G A 64 B 0 1 F byte byte 000 000 255 013 010 Start Calibration CLR CR 067 076 082 033 013 082 013 013 Calibration pH 2 111 CR 5 001 001 001 033 013 001 013 010 2 Calibration pH 4 112 CR 5 001 ...

Page 8: ...2 013 013 The AD converter or USB interface module is now ready to receive the calibration pH 7 command Initiate the pH7 calibration by sending the calibration pH 7 command 113 CR Allow some time for the module to stabilize this may take up to 2 minutes maximum Wait for the module to stabilize receive bytes 003 013 010 End the calibration process send QIT CR Wait for the module to confirm calibrat...

Page 9: ...he module to stabilize receive bytes 002 013 010 Rinse the probe with deionized water Place ISFET sensor and reference in the next calibration buffer solution pH 7 Initiate the pH7 calibration by sending the calibration pH 7 command bytes 113 CR Allow some time for the module to stabilize this may take up to 2 minutes maximum Wait for the module to stabilize receive bytes 003 013 010 Rinse the pro...

Page 10: ...ition A B C D E F G H I J K Received bytes 001 023 027 000 00 000 000 000 000 013 010 Protocol A 001 B 023 C 027 DEFGHIJK n a A 4096 B 64 C 1 4096 23 64 27 5595 pH value 5 595 5 2 4 Retrieving the temperature signal Retrieve temperature value Send the command bytes 777 CR Wait for the module to return the pH value receive bytes ABCDEFG Data ABCDEFG marks the byte position Values in bytes are neede...

Page 11: ... replace the ISFET sensor and or reference electrode Retrieve slope value Send the command bytes 000 CR Wait for the module to return the slope values receive bytes ABCDEFGHIJKLMN Data ABCDEFGHIJKLMN marks the byte position Values in bytes are needed for decoding Byte position A B C D E F G H I J K L M N Byte value 001 byte byte 002 byte byte 003 byte byte 004 byte byte 013 010 Decode received seq...

Page 12: ...12 5 3 ASCII table ...

Reviews: