Infrared Temperature Switch (Hot Metal Detector) Polaris PS09 / PI16
Configuring via SensorTools Software
14
Port test:
Enables testing of the outputs by simulating a switching temperature.
The simulation automatically stops after 3 seconds.
Interface:
For setting the
baud rate
(see
4.6.4
),
device address
(see
4.6.5
)
and
interface delay
(see
4.6.6
).
Service functions
To back up, for printing or restore device parameters or create service files for
remote diagnosis of problems.
- Create configuration file:
A configuration file includes all device settings made by user. This can
be used e.g. to setup a new device with the same values
and settings or to save the current status in case of a de-
vice replacement.
- Load Configuration file:
load a previously saved configu-
ration data in the device.
- Print parameters:
Prints the parameters of the currently
connected device.
-
Print configuration file: Prints a saved file.
- Create service file:
A service file includes all device data
and all software settings made by user. Creating such a file
can be useful if a problem can’t be solved and help from
our service could be useful.
- Upload service file:
Transmitting a (possibly modified) file
to the device it was created from.
Spot size calculator
The spot size calculator (from the
main window) is used to calculate
the spot size diameter in front and
behind the focused measuring dis-
tance (see
4.5.1
). To do this, fill the
white fields with the basic data (see
4.5
spot size tables
), then values
in the gray fields can be calculated.
Start simulation
Specify simulation
temperature
Check output state