- 3 -
(
)
2.2 Environmental
Requirements
Instrument
Followinglocation shall be recommendable for installations ;
Well ventilated
①
Ambient temperature shoud befrom -5
to 40
.
℃
℃
Cool and n o - heated
②
Alocation where is not exposedto direct sunshine, tohighradiantheat.
Less humid
③
:Donotinstall CE-8Eattheplacesofhighhumidityorundertheconditionsthat
Caution
thewaterandchemicalsmaysplashon.Thesemay becauses of the damages to
insulationorofrust.
Ambient humidity shoud be lower than 85 % RH.
Less mechanical vibrations
④
Maintenance andInspection works are easy
⑤
Others;
⑥
DonotinstallCE-8E attheplacewhereanycorrosivegasesordustsstore.
Avoidthe places whereare muchinfluenced to theelectromagnetic field.
(
)
2.3 Installation Instrument
Mountingmethodof the instrumentisforthepanel-mounttype.
①
Prepare the panel cut a s under mentioned.
For the application of multiple-linedpanel,clearances totheneighborpanelcutsshoudbe
kept horizontally 25mm or over, vertically 35mm or over.
Recommendable thickness of the panelshallbe 1~5mm.