- 2 3 -
5.3.6
Cable Length Correction Menu
Usingstandard sensor 10m , it is not necessarytoenterthismenu.
(
)
(
)
from Temperature Calibration Mode
Cable Length Correction Menu.
①
Push
key toenter
SEL
Cable Length Correction Menu.
Main display shows
to indicate
Secondary display shows c
.
able length total
(
)
Push
key.
②
ENT
Secondary display starts flashing to indicatechangeable value.
Push and release
keys to change the value to using cable length.
③
▲
▼
「
」
Onepushcountsup/downstepby step. And keeppushing,
counting up/downspeeds up.
(
)
Right fig. is example of 20m
Push
key after setting desired value.
④
ENT
Secondary display flashes rapidly andshowsitisdefined.
Display returns to the first step of this menu.
⑤
Now, this correction menu is completed.