TESTS AND CALIBRATION
SMART-CAL
(Methane Calibration Only)
From the working display access the menu by pressing and holding the
MENU button (B) until the top line of the display reads USER MENU.
Press and release the SAVE/ZERO button (C) until the bottom line
displays SMART CAL.
Place the instrument into the cradle provided on the left side of the Smart-
Cal Calibration Station. Attach the tubing from the station to the instrument
sensor.
Press the MENU button (B). The display will show SMART CAL
Communicating. Select the test from the Smart-Cal Station to be
performed. At the end of the test the instrument will beep 3 times and
display PASS or FAIL.
Retry the test if necessary by pressing the proper button on the Smart-Cal
Station again.
Press and release the POWER/MUTE button (A) to return the working
display. Remove the tubing and return instrument to service or send
instrument to the proper place for repair per company procedures.