a. If
PASSED is shown, the calibration was successful, and the calibration due date will be
automatically reset for that sensor. Remove the calibration gas.
b. If
FAILED is shown, leave the calibration gas attached and attempt calibration again.
If successful calibration cannot be achieved, remove the instrument from service and
contact the factory for assistance.
10. The display will return to the
CALIBRATION menu. Press and release the
A
button once to
return to the
USER MENU, or twice to return to the working display.
TO MANUALLY CALIBRATE HYDROGEN SULFIDE (H2S) 25 PPM
1. From the working display, press and hold the
B
button until
USER MENU is shown near
the top of the display.
2. Press and release the
C
button until
CALIBRATION is shown.
3. Press and release the
B
button to enter the
CALIBRATION menu.
4. Press and release the
C
button until
H2S 25 PPM is shown.
5. Attach
H2S 25 PPM
calibration gas to the end of the gooseneck. Ensure gas is flowing.
6. Press and release the
B
button to start the calibration.
7. Wait for either a
PASSED or FAILED result to be shown.
a. If
PASSED is shown, the calibration was successful, and the calibration due date will be
automatically reset for that sensor. Remove the calibration gas.
b. If
FAILED is shown, leave the calibration gas attached and attempt calibration again.
If successful calibration cannot be achieved, remove the instrument from service and
contact the factory for assistance.
8. The display will return to the
CALIBRATION menu. Press and release the
A
button once to
return to the
USER MENU, or twice to return to the working display.
TO MANUALLY CALIBRATE HYDROGEN CYANIDE (HCN) 10 PPM
1. From the working display, press and hold the
B
button until
USER MENU is shown near
the top of the display.
2. Press and release the
C
button until
CALIBRATION is shown.
3. Press and release the
B
button to enter the
CALIBRATION menu.
4. Press and release the
C
button until
HCN 10 PPM is shown.
5. Attach
HCN 10 PPM
calibration gas to the end of the gooseneck. Ensure gas is flowing.
6. Press and release the
B
button to start the calibration.
7. Wait for either a
PASSED or FAILED result to be shown.
a. If
PASSED is shown, the calibration was successful, and the calibration due date will be
automatically reset for that sensor. Remove the calibration gas.
b. If
FAILED is shown, leave the calibration gas attached and attempt calibration again.
If successful calibration cannot be achieved, remove the instrument from service and
contact the factory for assistance.
8. The display will return to the
CALIBRATION menu. Press and release the
A
button once to
return to the
USER MENU, or twice to return to the working display.
TO MANUALLY CALIBRATE CARBON MONOXIDE (CO) 100 PPM
1. From the working display, press and hold the
B
button until
USER MENU is shown near