![SENSIT Technologies GOLD G3 Instruction Manual Download Page 29](http://html1.mh-extra.com/html/sensit-technologies/gold-g3/gold-g3_instruction-manual_1243775029.webp)
29
USER MENU
The
SENSIT GOLD G3
has a user-accessible menu that is used for user-level tasks such as
changing the combustible gas type, performing a manual calibration, viewing logs, etc. A brief
description of each menu option is listed below, followed by more details for each in the following
section(s). Any options that are not available by default are listed as such (for example, a log for a
non-default test mode).
Other non-user adjustable settings are in the password protected Expert Menu.
To access the
USER MENU from the working display, press and hold the
B
button until
USER
MENU is shown near the top of the display. While in this menu, pressing and releasing:
• the
A
button will exit to the working display
• the
B
button will select the currently shown option
• the
C
button will scroll through each menu option
SET TIME
Allows the user to adjust the instrument time in 24-hour format (hours and
minutes). When installed and configured, this is set automatically by GPS.
SET DATE
Allows the user to adjust the instrument date (day, month, and year). When
installed and configured, this is set automatically by GPS.
TIME ZONE
Allows the user to set the instrument time zone (hours and minutes). This is
an offset from UTC +0.
GPS
OPTIONAL FEATURE
When installed and configured, this shows the status of the GPS signal
received by the instrument and the latitude and longitude coordinates.
FLASHLIGHT
Allows the user to turn the flashlight at the end of the flexible wand on or off.
BUMP TEST
Perform automatic test for sensor response to calibration gas within 60
seconds or less. A response of 80% or greater of the calibration gas value
indicates a successful test.
SMART SENSOR
Displays information pertaining to any smart sensors installed, including
sensor installation date and months of service.
CALIBRATION
Contains options for AUTOCAL and manual calibration of all sensors installed.
O2 TEST
Response test specific to the oxygen sensor, when installed, that checks
proper depletion of the O2 sensor when exposed to gas void of oxygen
(such as 100%VOL Methane calibration gas).
GAS TYPE
When the product is released, this will allow the user to switch the
combustible gas type. The only currently supported gas type is
NATURAL
GAS and no adjustment is currently allowable.
CAL LOG
Displays historical calibration information.
SESSION LOG
Displays manually saved gas readings.
BARHOLE LOG
OPTIONAL FEATURE
Displays automatically saved Barhole records.