background image

18

Gilian

®

 LAB & FIELD CALIBRATOR SYSTEM

Sensidyne Document No. F-PRO-1260 (Rev C)

3.3.2

Low Flow Control System Calibrations

This procedure applies to HFS-113 & HFS-513 sam-
plers with low flow capability and GilAir/GilAir-5 sam-
plers with Multi-Flow Modules in the “On” Position.

The suction/pressure regulator in conjunction with a
manifold provides the means for setting low flow

To determine or check these pressures, set sampler
flow rate, (between 1–1.5 LPM), blocking the pump in-
let and setting a back pressure of 20–21” H

2

O for HFS-

113 & HFS-513 pumps and 16” H

2

O for GilAir/GilAir-5

pumps, indication on the pressure gauge.

In this set up, the pump is connected to inlet (B4) and
the Load Selector Valve (V1) is set to “Bypass”.

Block the Air Inlet Fitting (B1). The flow rate will drop
to zero and the pressure gauge will peg at 20-21” H

2

O.

This can also be accomplished by closing (clockwise)
either the “High Flow Adjust” (V3) or “Low Flow Ad-
just” (V4) and then positioning the “Load Selector
Valve” to either position.

The GilAir with a constant-pressure module can be
checked in the same manner except there is no adjust-
ment for the suction/pressure, only verification.

The suction/pressure should be greater than 16” H

2

O.

To determine the stability of the regulator in all of the
above cases, set the Load Selector Valve (V1) to “By-
pass”, shut the Low Flow Adjust Valve (V4) by turning
fully clockwise, then set the Flowmeter Selector Valve
(V2) to the High flow range flowtube and connect the
sampler to the Air Inlet Fitting (B1).

Turn sampler ON and adjust the flow to 1 LPM HFS-113
& GilAir and 1.5 LPM for the HFS-513.

Turn the load selector valve to the “Low Flow Adjust”
position (V4). Be sure valve is shut off.

The regulator should go up to approximately 20-21”
H

2

O for HFS-113, 24-25” H

2

O for HFS-513 and a mini-

mum of 16” H

2

O for GilAir. If the pressures are out of

specifications, the pump should be leak checked and
the “Low Flow Adjust” reset to apply the proper pres-
sure range.

NOTE

It is imperative that flow rates are in the approximate
range, and flowtubes must be calibrated to determine
this.

Summary of Contents for Gilian 800565 Series

Page 1: ...________ HIGH FLOW ADJ V3 LOAD SELECTOR VALVE V1 BYPASS LOW FLOW ADJ V4 B4 B1 FLOWMETER SELECTOR VALVE V2 FM1 FM2 FM3 20 10 0 0 20 40 60 80 100 30 40 INCHES OF WATER cm of water MAX PRESSURE 15 PSIG D...

Page 2: ......

Page 3: ...el with 3 rotameter flow tubes Carrying Case Sampling Pump Gilibrator 2 Precision Flowmeter Not included with Calibrator Panel only model PN 800573 3 or Lab Stand Calibrator PN 800873 3 ALWAYS check t...

Page 4: ...e Gilian logo are registered trademarks These trademarks are protected through use and registration in the United States The trademarks and servicemarks used in this document are the property of their...

Page 5: ...SECTION TWO SYSTEM SET UP 2 1 Leak Testing 11 2 2 Flowtube Calibration 12 SECTION THREE PUMP SET UP OPERATION 3 1 Pump Performance Check 14 3 2 Pump Checkout 15 3 2 1 Load Simulation 15 3 2 2 Pressur...

Page 6: ...st 20 4 2 Flowtube Cleaning 23 APPENDICES Appendix A Parts List 24 25 Appendix B Returned Material Authorization 26 LIST OF FIGURES 1 1 Lab Field Calibrator Panel Front View 10 2 1 Calibration Plot Ex...

Page 7: ...dent DO NOTremove or alter any label or tag on this product its accessories or related products DO NOToperate this product should it malfunction or require repair DO NOTattempt to repair or modify the...

Page 8: ...Addi tional flowtubes are available to customize the calibra tor flowtube range to meet the users needs Specifications Flow Rates variable in 5 steps 0 5 5000 cc min 0 4 4000 cc min 50 800 cc min 30...

Page 9: ...acing the actual load i e the sampling media downstream of the flowtube which then gives direct pressure drop read ings and flow information about the loading source 1 3 DIAGNOSTIC PANEL TYPES 1 3 1 S...

Page 10: ...R SELECTOR VALVE V2 FM1 FM2 FM3 20 10 0 0 20 40 60 80 100 30 40 INCHES OF WATER cm of water MAX PRESSURE 15 PSIG DWYER INSTRUMENTS INC MICH CITY IND 46360 U S A PAT 3 397 319 3 862 416 4 011 759 4 347...

Page 11: ...the leak If leakage is detected set V2 to FM2 and repeat the entire procedure If no leakage is found it means the total system is intact and that the leakage found in the previous step lies in the tub...

Page 12: ...wmeter Selector Valve V2 10 to the flowtube being calibrated and connect the suction side of the Gilibrator 2 or bubble burette to the Air Inlet Fitting B1 7 3 Set the pump flow rate to 1000 cc min by...

Page 13: ...No F PRO 1260 Rev C Figure 2 1 Calibration Plot Example 5 4 3 2 1 1 2 3 4 5 ACTUAL FLOW IN LPM GILIBRATOR 2 OBSERVED FLOW IN LPM ROTAMETER 1 0 0 0 0 f t E l e v a t i o n a t 6 0 F S t a n d a r d C...

Page 14: ...ing the valve counterclock wise Note that this action will have no effect at this time 5 Switch the Load Selector Valve V1 to the Low Flow Adjust position V4 6 To simulate a 20 H2O filter loading back...

Page 15: ...1 9 LPM Continue to increase the back pressure on the pump by turning the High Flow Adjust Valve V3 clockwise until the flowtube reads 1900 cc min Wait 2 minutes If the pump does not fault increase th...

Page 16: ...Use only with UL Listed Portable Air Sampling Pump Battery Pack LFS 113 PUMP GILIBRATOR OR GILIBRATOR 2 PRECISION FLOWMETER UNIVERSAL CALIBRATION PANEL 5 0 4 0 3 0 2 0 1 0 0 5 200 150 100 50 40 30 20...

Page 17: ...e Back pressure Compensation Sys tem BCS in the pumps These potentiometers are adjusted at 2 flow rate points at 1 LPM for HFS 113 GilAir 1 5 LPM for HFS 513 at 3 LPM for HFS 113 GilAir 4 LPM for HFS...

Page 18: ...nd then positioning the Load Selector Valve to either position The GilAir with a constant pressure module can be checked in the same manner except there is no adjust ment for the suction pressure only...

Page 19: ...ontinue to close V4 to approximately 30 35 H2O or until the Fault light stays On Once the Fault light is On do not adjust the valve any further The pump should stop within one 1 minute after the red l...

Page 20: ...raduates are facing front 5 Slide three 3 pieces of tubing 9 onto upper manifold 19 and install onto the flowtubes at top of flowtube housing and secure with two screws 20 6 Rotate the selector valve...

Page 21: ...ECTOR VALVE V2 FM1 FM2 FM3 20 10 0 0 20 40 60 80 100 30 40 INCHES OF WATER cm of water MAX PRESSURE 15 PSIG DWYER INSTRUMENTS INC MICH CITY IND 46360 U S A PAT 3 397 319 3 862 416 4 011 759 4 347 744...

Page 22: ...22 Gilian LAB FIELD CALIBRATOR SYSTEM Sensidyne Document No F PRO 1260 Rev C Figure 4 2 Calibrator Panel Rear View...

Page 23: ...sheet metal flowtube housing Each individual flowtube can then be separated from the lower manifold 6 Clean the flowtube using lab grade alcohol in a sy ringe through the bottom 7 After cleaning blow...

Page 24: ...o c A s e b u t w o l f e g n a r w o l d n a h g i h h t i w e m o c s r o t a r b i l a C l e n a P s t n e m e r i u q e r c i f i c e p s r o f e l b a l i a v a s i s e s a C t n e m u r t s n I...

Page 25: ...l p e R 2 0 3 1 0 4 M P L 4 4 0 e b u T w o l F t n e m e c a l p e R 9 4 2 1 0 4 M P L 5 5 0 e b u T w o l F t n e m e c a l p e R s r e t e m a t o R g n i d n a t S e e r F o N t r a P n o i t p i...

Page 26: ...in advance for assistance with a problem which cannot be remedied and or requires the return of the product to the factory All returned products re quire a Returned Material Authorization RMA num ber...

Page 27: ......

Page 28: ...1000 112th Circle N Suite 100 St Petersburg Florida 33716 800 451 9444 727 530 3602 727 539 0550 FAX www sensidyne com Revision C Document No F PRO 1260...

Reviews: