Sensa Core Hemo Spark User Manual Download Page 4

Section 1     Introduction

Intended Use 

The  Hemo  Spark  Blood  Hemoglobin(Hb)  Measuring  System  is  intended  for  the  quantitative 
determination of hemoglobin (Hb) in capillary, venous & arterial human whole blood . The 

 testing system is designed for 

in vitro

 diagnostic use only (external use only).

The Hemo Spark Blood Hemoglobin Measuring System is for professional in vitro diagnostic 

 use only. Simply add a drop of blood to the test strip, and the result is 

 displayed on the 

 

screen  in  less  than  30  seconds.  This  device  has  not  been  evaluated  for 

pediatric  subjects.

 

4

* Read instructions carefully. Complete any necessary training before use.

* Only use Hemo Spark Hemoglobin (Hb) Test Strips with the Hemo Spark Hemoglobin (Hb) Meter.

* For Professional 

in vitro

 diagnostic use only.

* Test only whole blood specimens. Blood samples with EDTA or heparin anticoagulants can be used.

* This device has not been evaluated for pediatric subjects.

* Keep out of reach of children.

* For any queries, please contact Customer Support at 1800 212 2880

The Hemoglobin Test strip includes a mesh covered sample reaction zone. Sample is applied 
to the center of the sample reaction zone. The mesh functions to separate the sample evenly 
on the entire reaction layer. The reagents on the reagent layer function to hemolyze and 
release the hemoglobin. The hemoglobin is converted to methemoglobin to cause a color  

Principle of Operation for the Methodology

The meter reads the reflection of the strip at every second until the end point of the reaction 
is  detected.  The  reflection  at  the  end  point  is  directly  proportional  to  the  hemoglobin 
concentration. The end point is defined as following: 

To ensure accurate results:

change on the strip.

The reflection changes between ±1% in three continuous second. Then the reflection at the 

last second will be read as the end point. 

The Hemo Spark Blood Hemoglobin(Hb) Measuring System provides results in less than 
30 seconds.  The test  only  requires  10 μL drop  of whole  blood.  The meter  can store  up to 
1,000 results and can be powered by 3 AAA rechargeable batteries (Lithium/NiMH

/Zinc

/ZnCl

  Carbon) or an optional AC adapter.

Summary of Contents for Hemo Spark

Page 1: ...g dL n WARRANTY WARRANTY YEARS YEARS 3 3 3 YEARS Blood Hemoglobin Measuring System ISO 9001 ISO 13485 Obelis s a Bd General Wahis 53 1030 Brussels Belgium Tel 32 2 732 59 54 Fax 32 2 732 60 03 mail ob...

Page 2: ......

Page 3: ...p and Options 16 Test Number Setup 17 Meter Setup 18 Testing 20 Instructions for use of Lancing Device 21 Instructions for use of Safety Lancets 24 Data Transfer Transmission 28 Deleting Data 28 Optic...

Page 4: ...eries please contact Customer Support at 1800 212 2880 The Hemoglobin Test strip includes a mesh covered sample reaction zone Sample is applied to the center of the sample reaction zone The mesh funct...

Page 5: ...eins Hemo Spark Blood Hemoglobin Hb Measuring System bones and tendons Suggested Test Areas for the Hand Alternate Test Sites Warning We recommend that you test on your finger tips for better results...

Page 6: ...suring System Section 2 Contents in Hemo Spark Kit Hemo Spark kit will have the items as per CONTENTS mentioned on the kit box Before testing read the instructions carefully and learn about all the co...

Page 7: ...in Hb concentration and hematocrit Hct value Below diagram describes the parts of your meter Meter 1 LCD Liquid Crystal Display 2 Left Arrow Button 4 Strip Channel 3 Setup Button 5 Right Arrow Button...

Page 8: ...Indicates test result or displays menu options Memory Indicates a test result is being recalled from memory Measurement Units Indicates the units for the test result Hct Value Shows calculated Hct va...

Page 9: ...l local regulations when disposing of the meter and used batteries All Hb Systems Preventive Warnings with Regard to EMC This instrument is tested for immunity to electrostatic discharge as speci ed i...

Page 10: ...capillary and venous whole blood or arterial whole blood 1 Insert Arrows Located on the front of the test strip These arrows indicate the direction in which the test strip should be inserted 2 Sample...

Page 11: ...f there is too much specimen covering the area repeat the test with a new test strip Note Do not add more blood to the test strip if the sample applied to the sample application area When a Lo or E 5...

Page 12: ...ions The Hemo Spark Hemoglobin Hb Meter Control Strip is a plastic strip containing a brown reference pad which works with the Hemo Spark Hemoglobin Hb Meter to ensure the optical system is working pr...

Page 13: ...d direct sunlight Transport and store the control strips in its closed vial within 2 40 C 35 6 104 F Do not freeze or refrigerate Replace the cap on the control strips vial immediately after removing...

Page 14: ...or into a 100 240V AC 50 60Hz primary power outlet Note If the meter is having batteries they will recharge automatically while the meter is connected with the AC Adapter power supply The meter can al...

Page 15: ...label Results may be inaccurate if the two numbers are not identical If the code number on the code chip does not match the number on the test strip vial with which it was packaged Please contact your...

Page 16: ...modes Press or to select and press button to enter following setup sub modes No SEt Test number setup The test number can be from 1 to 999 Refer to Page 17 CHE Optical Check mode Refer to Section 8 SY...

Page 17: ...r Setup mode Test Number Setup The test number can be set to any number from 1 999 Press or until the correct test number is displayed Press to save and return to No SEt screen Note Once the meter rea...

Page 18: ...ear ashing and y displayed on test result area Then press button to save and proceed to setup time Press or until the SYS is displayed From the SEt screen Press to enter Meter Meter Setup The rst opti...

Page 19: ...eset to 1 for each new day of testing when the test number reset is turned on Press to save and proceed to setup unit OFF means test number will continue the last number when meter turns o at last tim...

Page 20: ...s Mix the specimen well Then collect approximately 10 L into a capillary transfer tube Apply it to the center hole of the sample application area of the strip Do not touch the test strip with the tube...

Page 21: ...21 Instructions for use of Lancing Device...

Page 22: ...22...

Page 23: ...23...

Page 24: ...s will help to prevent from infections Discard the used safety lancet carefully after each use to avoid untended safety lancet stick injuries Used safety lancets maybe considered bio hazardous waste i...

Page 25: ...tup to ensure all segments and display elements are turned on and there are no missing icons or elements The meter will brie y show a blank display Observe there are no segments or icons permanently t...

Page 26: ...area of the test strip Note For testing capillary blood use the second drop of blood for accurate results Refer to Section 6 for Fingertip Testing details Insert a test strip into the Strip Device Ch...

Page 27: ...mens and materials Perform daily cleaning when testing is completed for the day Refer to Section 9 Maintenance Themeterwillautomaticallyturno after5minutesofinactivity orwhen buttonispressed If the me...

Page 28: ...efer to Section 5 Meter Setup press or until PC is displayed Press button to enable the Data Communication mode MEM will be displayed Press button to transmit data to an external certi ed PC After dat...

Page 29: ...rn to the Initial Screen If no data is stored the meter will display one dash and MEM Press button to enable data deletion MEM will be displayed Press button to returns to the Setup Menu Memory Databa...

Page 30: ...104 F prior to testing The optical check should be performed under normal lab lighting conditions Do not perform under sunlight or extreme lighting conditions Note Press button to enter this mode The...

Page 31: ...utton to return to the Home Screen or Long Press or to return g dL the speci ed ranges of Quality Control levels then the meter is not functioning properly OR and the meter displays no Setup Menu Chec...

Page 32: ...the cover to open it Remove and discard the old batteries Insert new 3 AAA rechargeable batteries Lithium NiMH Zinc Carbon ZnCl Close the battery cover and make sure that it snaps shut Recheck and res...

Page 33: ...condition requirements Keep the meter clean Wipe it frequently with a soft clean and dry cloth Use fresh water when needed Do not clean the unit with substances such as gasoline paint thinner or other...

Page 34: ...recharge the batteries Insu cient sample Number of tests can be Code chip was removed Repeat test and apply enough sample Use Replace with new code chip If the code chip is removed during testing aro...

Page 35: ...ht 72 g without batteries Meter Storage conditions 0 50 C 32 122 F Operating Conditions 2 40 C 35 6 104 F Analyzer Port USB cable for Data Transfer or Power optional Capabilities Meter Warranty USB Da...

Page 36: ...Do not discard along with household waste USB Port Caution Keep away from sunlight and heat Do not reuse Keep Dry Do Not Use if package is Damaged Storage or Operating Temperature Manufacturing Date...

Page 37: ...rts required which were not defective shall be replaced at additional cost Sensa Core shall not be required to make any repairs or replace any parts that are necessitated by abuse accidents alteration...

Page 38: ......

Page 39: ...yderabad Telangana INDIA Tel 91 8455 223400 www sensacore com Sensa Core Medical Instrumentation Pvt Ltd SC PHSC 0422A Customer Care Details For any feedback complaints please contact our Customer car...