Section 8 Optical System Check
Press Set up button and
►
to find Check Control Strip mode.
* The control strip is intended for checking the optical system.
* Allow the control strip and the meter to reach temperature 2-40°C
(35.6 - 104ºF) prior to testing.
* The optical check should be performed under normal lab lighting conditions.
Do not perform under sunlight or extreme lighting conditions.
Note:
Press button to enter this mode. The meter will flash the strip symbol as shown below.
Insert a control strip into the Strip Channel in the same direction as the arrows indicate on
the strip or device. Ensure that the control strip is inserted all the way.
30