background image

 

 

ORIGINAL INSTRUCTIONS  

 

USER MANUAL

 

Z-SG2 / Z-SG2-L 

 

DIGITAL LOAD CELL CONVERTER WITH 24 BIT ADC

 

 

 

 

 

 

 

 

SENECA S.r.l. 

Via Austria 26 – 35127 – Z.I. - PADOVA (PD) -  ITALY  

Tel. +39.049.8705355 – 8705355  Fax +39 049.8706287 

www.seneca.it 

Summary of Contents for WZ-SG2

Page 1: ...GINAL INSTRUCTIONS USER MANUAL Z SG2 Z SG2 L DIGITAL LOAD CELL CONVERTER WITH 24 BIT ADC SENECA S r l Via Austria 26 35127 Z I PADOVA PD ITALY Tel 39 049 8705355 8705355 Fax 39 049 8706287 www seneca...

Page 2: ...correct installation and the user must follow all the operations described in the installation manual carefully Seneca is not responsible for failures breakages and accidents caused by ignorance or fa...

Page 3: ...2019 1 0 0 0 First revision MM 16 09 2019 1 0 0 1 Added info for bootloader mode Updated info for the USB port Added a note for the Auto Resolution Added A Resolution Warning MM 26 11 2019 1 0 0 2 Fix...

Page 4: ...15 8 ANALOGUE OUTPUT CONNECTIONS Z SG2 ONLY 15 8 1 1 RS485 PORT CONNECTIONS 16 8 2 CONFIGURATION OF THE RS485 PORT BY DIP SWITCH 18 9 CONFIGURATION OF THE FLASH MEMORY RS485 PORT 18 9 1 SETTING THE B...

Page 5: ...ELL FULL SCALE 27 10 2 13 2 EXCEEDING THE THRESHOLD AND STABLE WEIGHING 27 10 2 13 3 STABLE WEIGHING 28 10 2 13 4 MODBUS CONTROLLABLE 28 10 2 13 5 EXCEEDED THRESHOLD WITH HYSTERESIS ALARM 28 STABLE WE...

Page 6: ...ORY CALIBRATION WITH MANUALLY INSERTED 14 2 TARE 37 CALIBRATION PROCEDURE WITH AN EXTERNAL DEVICE CALIBRATION WITH SAMPLE WEIGHT 37 14 3 MODBUS RTU SLAVE COMMUNICATION PROTOCOL 38 15 SUPPORTED MODBUS...

Page 7: ...the functions fully meet the customer s expectations or that the device firmware and software should have no errors or operate continuously DESCRIPTION 1 1 Z SG2 is a load cell converter strain gauge...

Page 8: ...al input output Yes configurable between input output USB port Yes on the front for the configuration calibration register reading Drivers Virtual com port for Windows Linux MAC OS Windows CE Configur...

Page 9: ...5200 bit s configurable Parity Delay in replying Configurable USB RS485 port protocol Modbus RTU Slave LED SIGNALS 2 Z SG2 Z SG2 L LED 2 1 LED SIGNALS ON THE Z SG2 Z SG2 L FRONT LED STATUS LED MEANING...

Page 10: ...immunity requirements use shielded signal cables connect the shield to a preferential instrumentation earth system separate shielded cables from other cables used for power installations transformers...

Page 11: ...y 10 40 C Since the cable resistance depends on the temperature the response of the cable to temperature changes must be eliminated The 4 wire cable is part of the load cell temperature compensation s...

Page 12: ...cell has 6 wires check that the same voltage is also measured between Sense and Sense Now leave the cell at rest without the tare and check that the voltage between the Signal and Signal cables is aro...

Page 13: ...ARALLEL IMPEDANCE OF THE STATED LOAD CELL Ohm MAXIMUM NUMBER OF CONNECTABLE CELLS IN PARALLEL 350 4 1000 8 For the connection of 4 load cells Seneca recommends using the SG EQ4 product To connect 2 or...

Page 14: ...20 potentiometer is inserted in the Excitation cable of each load cell There are two ways to trim the load cells The first method is to adjust the potentiometers by trial shifting the calibration wei...

Page 15: ...value of the other load cells 5 Measure and adjust the excitation voltage of the other three load cells using the respective potentiometer Check the results and make a final adjustment by moving a tes...

Page 16: ...RIOR PERMISSION www seneca it Doc MI 005370 EN Rev 6 Page 16 Analogue current output RS485 PORT CONNECTIONS 8 2 The device has an RS485 serial port to be connected to the Z PC DINAL accessory and ther...

Page 17: ...SG2 Z SG2 L ALL RIGHTS RESERVED NO PART OF THIS PUBLICATION MAY BE REPRODUCED WITHOUT PRIOR PERMISSION www seneca it Doc MI 005370 EN Rev 6 Page 17 The protocol available through the RS485 port is Mod...

Page 18: ...address of the RS485 port ATTENTION The configuration via dip switch is active only after a restart ATTENTION The configuration via dip switch has priority over the one saved in flash if you need to u...

Page 19: ...Page 19 SETTING THE BAUD RATE OF THE RS485 PORT BY DIP SWITCH 9 2 Dip switches 1 and 2 configure the baud rate Baud Rate DIP1 DIP2 9600 OFF OFF 19200 OFF ON 38400 ON OFF 57600 ON ON ATTENTION The pari...

Page 20: ...y convention Modbus Station Address DIP3 DIP4 DIP5 DIP6 DIP7 DIP8 1 OFF OFF OFF OFF OFF ON 2 OFF OFF OFF OFF ON OFF 3 OFF OFF OFF OFF ON ON 4 OFF OFF OFF ON OFF OFF 5 OFF OFF OFF ON OFF ON 6 OFF OFF O...

Page 21: ...TOR BY DIP 2 OF SW3 9 5 Usually it is not necessary to use a terminator on the RS485 line unless this is inserted into the bus by some other device for example some modbus masters have a terminator th...

Page 22: ...70 EN Rev 6 Page 22 Send Sends the configuration if the dip switches ARE NOT ALL OFF the device uses the configuration of the RS485 by dip switch and NOT the one sent by the configuration Test and Cal...

Page 23: ...dip switch settings OPERATING MODE 10 2 1 It allows to configure the basic operation of the device can be set to factory calibration or to Calibration with Sample weight 10 2 1 1 FACTORY CALIBRATION...

Page 24: ...urement will have the sign if traction it will have the sign A typical case of use is to link the direction of the force to the analogue output so that for example 4mA correspond to the maximum tracti...

Page 25: ...s to be used only by experts or on the advice of Seneca personnel ADVANCED FILTERING 10 2 9 These parameters can only be configured by expert personnel or at the suggestion of Seneca 10 2 9 1 NUMBER O...

Page 26: ...t is in rough ADC points 10 2 9 5 FILTER RESPONSE SPEED It represents a parameter related to the filter response speed it can vary from 0 001 slowest response to 1 fastest response Represents the vari...

Page 27: ...nnel is configured as an input or an output DIGITAL INPUT CONFIGURATION 10 2 12 If the digital channel is set as an input it is possible to select its behaviour 10 2 12 1 ACQUIRES THE TARE In this mod...

Page 28: ...ontrolled directly by modbus see STATUS register in the chapter 16 6 10 2 13 5 EXCEEDED THRESHOLD WITH HYSTERESIS ALARM In this mode the output is activated when the net weight reaches the threshold t...

Page 29: ...ET WEIGHING ALARM 10 2 15 It is the threshold value in technical units see 10 2 13 HYSTERESIS 10 2 16 It is the hysteresis value used in the digital output mode EXCEEDED THRESHOLD WITH HYSTERESIS see...

Page 30: ...ts the baud rate to be used in the communication from the RS485 port PARITY 10 2 21 It sets whether the communication of the RS485 port must take place with or without even or odd parity DELAY IN REPL...

Page 31: ...MEASUREMENTS 10 2 25 It allows to configure the type of representation of the floating point values of the measurements while the configuration parameters do not change operation 10 2 25 1 FIRST PART...

Page 32: ...crosoft Excel The sample acquisition time can be set minimum 1 second The data logger will create a text file in a standard csv format CALIBRATION OF THE LOAD CELL WITH THE EASY SETUP SOFTWARE 11 In t...

Page 33: ...In the tare acquisition mode from the field proceed as follows for calibration 1 Replace the tare on the cell 2 Wait for the measurement to stabilize 3 Press the TARE ACQUISITION FLASH button FACTORY...

Page 34: ...ion 1 Set the Tare value in technical units 2 Press the SET MANUAL TARE FLASH button ATTENTION To obtain better measurement accuracy use the tare acquisition mode from the field CELL CALIBRATION WITH...

Page 35: ...l 2 Wait for the measurement to stabilize 3 Press the TARE ACQUISITION FLASH button 4 Replace the Tare Sample Weight 5 Wait for the measurement to stabilize 6 Press the SAMPLE WEIGHT ACQUISITION FLASH...

Page 36: ...nloaded from the device s web page The drivers are however installed automatically with the installation of the Easy Setup software DRIVERS FOR LINUX SYSTEMS 12 2 In many Linux distributions the drive...

Page 37: ...calibration of a load cell by sending Modbus commands depending on the type of operation chosen CALIBRATION PROCEDURE WITH AN EXTERNAL DEVICE FACTORY CALIBRATION WITH 14 1 TARE ACQUIRED FROM THE FIELD...

Page 38: ...EGISTER 9 The device saves the new sample weight value in flash and resets the COMMAND REGISTER value 10 The load cell is calibrated MODBUS RTU SLAVE COMMUNICATION PROTOCOL 15 The communication protoc...

Page 39: ...User Manual Z SG2 Z SG2 L ALL RIGHTS RESERVED NO PART OF THIS PUBLICATION MAY BE REPRODUCED WITHOUT PRIOR PERMISSION www seneca it Doc MI 005370 EN Rev 6 Page 39...

Page 40: ...d only RW Read write register RW Reading and writing register contained in flash memory writable a maximum of 10000 times Unsigned 16 bit unsigned integer register can take values from 0 to 65535 Sign...

Page 41: ...S OFFSET NUMBERING OF MODBUS ADDRESSES WITH 1 BASED CONVENTION STANDARD 16 1 2 The numbering is that established by the Modbus consortium and is of the type HOLDING REGISTER MODBUS ADDRESS 4x MEANING...

Page 42: ...0 1 1 0 0 0 0 0 0 0 0 1 1 0 0 MSB and LSB BYTE CONVENTION WITHIN A MODBUS HOLDING REGISTER 16 3 A Modbus Holding Register consists of 16 bits with the following convention BIT 15 BIT 14 BIT 13 BIT 12...

Page 43: ...the most significant word with the least significant word therefore it is possible to obtain 40064 as LSW and 40065 as MSW TYPE OF 32 BIT FLOATING POINT DATA IEEE 754 16 5 The IEEE 754 standard https...

Page 44: ...6 BIT 40002 1 FIRMWARE REVISION FIRMWARE REVISION RO UNSIGNED 16 BIT 40003 2 UNIT UNIPOLAR MSB UNIT Selects the unit of measurement among 0 Kg 1 g 2 t 3 lb 4 l 5 N 6 bar 7 atm 8 Other LSB UNIPOLAR 0 A...

Page 45: ...e in number of characters RW UNSIGNED 16 BIT 40006 5 OUTPUT STOP SCALE V mA Full scale value of the analogue output in V mA MSW RW FLOATING POINT 32 BIT 40007 6 Full scale value of the analogue output...

Page 46: ...40030 29 Ratio value of the load cell sensitivity in mV V MSW RW 40031 30 CELL FULL SCALE kg g t Full scale value of the load cell in technical units MSW RW FLOATING POINT 32 BIT 40032 31 Full scale v...

Page 47: ...HT kg g t Value of the delta weight in technical units for the stable weighing condition MSW RW FLOATING POINT 32 BIT 40042 41 Value of the delta weight in technical units for the stable weighing cond...

Page 48: ...anced is set RW UNSIGNED 16 BIT 40047 46 ADVANCED ADC CHOPPING Enables or disables the ADC Chopping mode For further information please refer to 10 2 9 2 It is active only if level 7 filtering advance...

Page 49: ...filtering advanced is set MSW RW FLOATING POINT 32 BIT 40052 51 Represents the variation in ADC points due to noise alone It is active only if level 7 filtering advanced is set LSW RW 40053 52 ADVANCE...

Page 50: ...ilter level FILTER LEVEL RESPONSE TIME ms 0 16 1 32 2 50 3 250 4 400 default 5 900 6 1800 7 ADVANCED Dependent on the configuration The higher the filter level the more stable but slow the weight meas...

Page 51: ...ion calculated based on full scale to get 10000 points DENOISE FILTER ENABLE LSB 0 Noise filter disabled advanced mode 1 Noise filter enabled RW UNSIGNED 16 BIT 40057 56 MANUAL RESOLUTION kg g t Sets...

Page 52: ...65535 In the case of bipolar measurement It can vary from 0 max traction to 32767 min traction From 32768 condition of 0 or rest to 65535 max compression RO UNSIGNED 16 40063 62 RESERVED RO UNSIGNED...

Page 53: ...R GROSS WEIGHT kg g t Value of the gross weight in technical units MSW RO SIGNED 32 BIT 40073 72 Value of the gross weight in technical units LSW RO 40074 73 INTEGER TARE WEIGHT kg g t Value of the ta...

Page 54: ...Stable weighing BIT 1 RO Bit 1 1 The cell has reached full scale BIT 2 RO Bit 2 1 Net weight is 0 BIT 3 RO Reserved BIT 4 RO Bit 4 1 Weighing is stable BIT 5 R W Only if the modbus controlled output m...

Page 55: ...e tare in RAM on restart it is lost 49914 decimal Acquires the tare in Flash for the calibration procedure in both operating modes factory calibration and with sample weight 50700 decimal Acquires the...

Page 56: ...s from restart LSW RO 40084 83 MIN NET WEIGHT kg g t Minimum value of the net weight in technical units from restart MSW RO FLOATING POINT 32 BIT 40085 84 Minimum value of the net weight in technical...

Reviews: