SP395 SoundPro Audio Integrator
Form7492
Operation Manual
99
3.
Enter code.
The code is an 8-character hexadecimal value. It contains the numbers 0-9
and the letters A-F. One at a time, select the character position, beneath the name of the
package, and click and turn the encoder to change the value to match the character in your
unlock code. Do this for all 8 characters.
4.
Exit function.
When you are done, exit the function to store the values. The function
should now be unlocked.
Setup - General
This menu has fields for miscellaneous data values that are stored in the SoundPro. Each is
described below. Since you can alter the contents of the calibration values stored in the
SoundPro, be careful with these data fields!
If you are unsure of what to do, contact Sencore
technical support before altering any setting
.
The General screen provides information regarding Burn-In, Factory Calibration and Restoring Calibration.
1.
Burn-in
-
This field indicates whether or not the SoundPro has been through the burn-in
process.
When a SoundPro is first powered up in manufacture, it enters a burn-in and self-calibrate
state. Several constants are stored at this time. The unit may be forced through the burn-in
process again by setting the burn-in data field on the General Setup menu to No, exiting
the menu, and powering the unit off and back on. The burn-in procedure takes one hour to
complete. Note that this may affect the main calibration.
2.
Factory Calibration
- This field indicates whether or not the SoundPro has been through
the factory calibration process. The factory calibration process requires a special
calibration fixture.
Contact Sencore technical support before changing this field.
3.
Restore Factory Calibration -
The factory calibration constants are stored in non-volatile
memory. A second copy of the constants is also stored so that, if the first set is altered, the
original factory calibration can be restored. To do this, set the Restore Cal data field on
the General Setup menu to Yes and exit the menu. The original factory calibration will be
restored.