J.P. SELECTA s.a.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
Tel (34) 93 770 08 77 Fax (34) 93 770 23 62
e-mail: [email protected] - http://www.jpselecta.es
GUIDE DE L’UTILISATEUR
CODE 80158.02
REV 0 02/10/09
(Sous réserve de modification)
Pag.: 12
S ºC MIN PR STA
1 1 2 5 3 0
1 STP
S ºC MIN PR STA
Select the
program
to check. Use keys
to check. Use keys
5. Checking program parameters:
The temperature / time parameters can be checked from the STP state.
Use keys
to move at
step
position on the display
S ºC MIN PR STA
1 125 30 1 STP
S ºC MIN PR STA
1 125 30 1 PRG
S ºC MIN PR STA
3 145 OFF 1 PRG
S ºC MIN PR STA
1 125 30 1 PRG
S ºC MIN PR STA
1 125 30 1 PRG
S ºC MIN PR STA
2 125 20 1 PRG
Select the
program
to edit. Use keys
(Default: program 1)
Select the
step
to edit. Use keys
(Default: step 1)
Set temperature for the step. Use keys
(Temperature must be always higher than the previous step)
(Default: previous step temperature)
Validate
Validate
Validate
Set time (minutes) for the step. Use keys
(Default: OFF)
•
•
•
Program
finish when there is an
step
with the
time
set at value OFF.
S ºC MIN PR STA
1 145 31 1 STP
Validate
Once,
a
step
with the
time
set at value OFF, is validated, it left pro-
gramming status and returns to STOP status.
4. Programming menu:
+ 5 seg.
Automatically jump to the next
step
.