INSTRUCTION MANUAL CODE 80188 REV E
Jun/2018
(It can be modified without notice)
Page: 38
J
.P. SELECTA s.a.u.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: [email protected] - website: http://www.jpselecta.es
Fig. 5.1 Control panel
6
1
3
2
7
8
4
5
9
10
PARAMETERS PROGRAMMING
Factory settings:
Low level water alarm:
36
% heating:
5
Distillation power:
80%
Volume of Boric acid:
4
Volume of HaOH:
12
5.3 Parameters settings
These settings must be done in both modes: «MANUAL» and «AUTOMATIC»
• Setting the distilling time
:
- Press
.
- Use keys
to set the distilling time (minutes).
• Setting the “Low water level alarm” time (low level of water into
steam generator)
- Press
.
- Keep pressed keys
for a 10s until display shows dashes.
- Press
to select steam.
- Use
to set the alarm steam time (factory setting 36).
- Use
to validate the setting.
- Press
to save the setting and exit from the settings mode.
• Setting the heating power on stand-by status:
- Press
.
- Keep pressed keys
for a 10s until display shows dashes.
- Press
to select % stand-by heating.
- Use
to set to set the % heating (factory setting 5).
- Use
to validate the setting.
- Press
to save the setting and exit from the settings mode.
• Setting distillation power:
- Press the keys
for 5 seconds until the display shows dashes.
- Press
and
until the display shows the current value.
- Modify the value by means of keys
.
- Press
and
to validate the setting.
- Press
to save and exit from the settings mode.