![Selecta 4002423 Manual Download Page 49](http://html1.mh-extra.com/html/selecta/4002423/4002423_manual_1236745049.webp)
MANUAL DE INSTRUCCIONES
CODIGO 80342 REV O
11/2021 Pag.: 49
J
.P. SELECTA s.a.u.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: [email protected] - website: http://www.grupo-selecta.com
To select a program, click on .
Press «BACK» to deselect the program.
Once the program has been selected, start it by clicking on .
The «Operation» screen appears.
Note: at the end of the list there is a program with fixed parameters for
cleaning the camera.
Sel
RUN
Number of the program
1 121.0
º
C 00h 03m +1.60bar
PRE 80
º
C
PUR 98
º
C 05m 00s
P03
Sel
back
2 80.0
º
C 00h 05m +1.60bar
RUN
Shows if program is selected or not
Program number
Shows if program is selected or not
1 121.0
º
C 00h 03m +1.60bar
PRE 80
º
C
PUR 98
º
C 05m 00s
P03
Sel
back
2 80.0
º
C 00h 05m +1.60bar
RUN
7.11 Operation screen
When you click on «RUN», the operation screen appears:
There are two operating screens:
Global view:
- It presents all the “steps” of the program.
- Indicates the current step.
- The water’s temperature.
- The temperature of the heart probe.
- The pressure.
- The current F
0
value.
- The total time.
- The function that is currently performing.
P03
hEART
117.0 ºC
wATER
121.3 ºC
PRESSURE
1.75 BAR
FO
1.7
global
view
Step
Stop
CRZN
CALENTANDO TIEMPO TOTAL: 00
h
46
M
23
S
1 121.0
º
C 00h 03m +1.60bar
PRE 80
º
C
PUR 98
º
C 05m 00s
2 80.0
º
C 00h 05m +1.60bar
Step view:
- Presents all the
current values
and
set
points of the current
step.
- It indicates the current step.
- The water’s temperature.
- The temperature of the heart probe.
- The pressure.
- The current F
0
value.
Jump to next STEP.
STOP operation.
It allows to modify the time set.
It allows to modify the pressure setpoint.
Low water level.
High water level.
Show GLOBAL VIEW
It shows the temperature
control probe.
Below low water level.
T. HEART
T. WATER
SET
67.3
º
C 34.5
º
C 121.0
º
C
CONTROL
TIME SET
01h 30m
01h 40m
PRSSURE SET
+0.03
bar
+1.50 bar
GLOBAL
VIEW
Step
Stop
F
0
SET
3.2 3.5