7.7 Parameter settings
Language
Press, successively on the box to select Spanish, English or French.
Time and Date
Press, successively on the frame to select the setting of day, month, year,
hour, minute, second.
Adjust by clicking on the icons increase and decrease.
Units
Press, successively on the frame to select the units of temperature and pres
-
sure.
Password
Reserved area where operating parameters of the autoclave are established.
The access password is 0000 for calibrator and 0284 for technical service.
For security of the function, it is recommended that this password to be known
only by qualified personnel.
Printer
Indicates whether the printer is installed or not.
Water
Always leave activated. Do not change.
Brightness
Regulates the brightness of the screen.
Change password
You can change the password.
F
0
Custom
The equipment incorporates three lethality tables of three reference mi-
croorganisms and a self-calculated lethality table so that the client can
incorporate the microorganism of their choice to calculate the total lethality
value of their heating treatment.
Lethality (L): Time (minutes) which it is necessary to heat a food product
to a certain temperature (T) in order to obtain the same lethal effect as
heating to the reference temperature (Tx) in 1 minute, singular value for
each microorganism.
The lethality values sum (L) during the time (minutes) the heat treatment
lasts will give us the total lethality value or the final F
0
sterilization or
pasteurization value.
Default lethality values (F
0
)
Default tables:
F
0
STERILE:
Lethality table of microorganism
Clostridium Botulinum
. This
is the reference microorganism (its sporulated form, more specifically) for
all thermal sterilization treatments.
F
0
PAST L:
Lethality table of microorganism
Streptococus faecalis D.
This
is a reference microorganism used in heating treatments of pasteurization
at low temperatures.
F
0
PAST H:
Lethality table of microorganism
Bacillus Coagulans
. This is
a reference microorganism used in heating pasteurization treatments at
high temperatures.
home
cal
LANGUAGE
DATE
UNITS
PASSWORD
INSERT
PRINTER
WATER
BRIGHT
PASSWORD
CHANGE
Consult with
any
questions before modifying a parameter
of the autoclave.
Downloading logs
When logs are downloaded, a copy of all logs is
downloaded to the USB Pendrive.
We suggest that you use the
registration
number
as the
lot number
.
Logs are removed from the autoclave, when
downloaded.
The maximum number of records that can re
-
main stored in the autoclave memory is about
100.
From this, they are deleted by order of antiquity.
The equipment will warn when the memory will
be above 80% full.
Download the logs every so often to store them
in a safe place.
We recommend keeping a copy in two different
USB memory sticks.
UNLOAD LOGS
INFORMATION
F0 CUSTOM
Incorrect modification of
operating parameters
can lead to incorrect treatments
or dangerous situations.