Manual revision 041
Section 4: Deploying and Operating MicroCAT
SBE 37-SM RS-232
43
Data Upload
Commands
Stop sampling (send
Stop
) before uploading data.
GetSamples:b,e
Upload data from scan
b
to scan
e
,
in format defined by
OutputFormat=
.
First sample is number 1. As data is
uploaded, screen first displays
start time =
start sample number =
These are start time and starting sample
number for requested data.
DDb,e
Upload data from scan
b
to scan
e
,
in alternate converted decimal form
(
OutputFormat=1
) (regardless of
OutputFormat=
).
First sample is number 1.
As data is uploaded, screen first displays
start time =,
start sample number =
.
These are start time and starting sample
number for requested data.
Notes:
Use Seaterm232’s Upload menu
to upload data that will be
processed by SBE Data
Processing
. Manually entering a
data upload command does not
produce data with the required
header information for processing by
our software. These commands are
included here for
reference for users
who are
writing their own software.
If not using the Upload menu
-
To save data to a file, click
Capture before entering a data
upload command.
See
Data Formats
after these
Command Descriptions
.
Example:
Upload samples 1 to 200 to a file (user input in bold).
(Click Capture menu and enter desired filename in dialog box)
GETSAMPLES:1,200
or
DD1,200