Section 4: Deploying and Operating MicroCAT
SBE 37-SM RS-232
56
The Configuration dialog box (which appears if you click
Modify
on
the File Setup tab) looks like this:
The settings in the .xmlcon file created by Seaterm232 are based on
the setup of the MicroCAT.
Review the deployment latitude, and modify as needed.
If your MicroCAT does not have a pressure sensor, review the
deployment pressure, and modify as needed.
Click Save if you made any changes, and then click Exit.
Time between scans. Must agree with
MicroCAT setup (
SampleInterval=
);
see reply from
GetCD
or
DS
.
Indicates if MicroCAT
includes pressure sensor.
If no pressure sensor
included, deployment
pressure is used to
calculate conductivity (and
derived variables such as
salinity and sound
velocity). Value shown is
based on
ReferencePressure=
that
was programmed into
MicroCAT;
you can
change this value in
.xmlcon file, if you have
updated deployment
depth information
.
Latitude is used to calculate local gravity (to
calculate salt water depth). If enabled,
software uses input latitude in calculation. If
disabled, software uses Latitude on
Miscellaneous tab of Data Conversion.
Enter latitude for your deployment
.
Indicates whether MicroCAT includes dissolved
oxygen sensor (IDO or ODO MicroCATs only).
Double click on sensor to view and/or modify
calibration coefficients, which are based on
calibration coefficients that were programmed
into MicroCAT.