2.0 Safety and Preparation for Use 

Page 2 

The  PPS-2RM-B  was  designed  for  indoor  use  only  and  is  not  intended  for  operation 
outdoors or in a wet environment. The instrument may be mounted in a standard 19-inch 
instrumentation  rack  or  may  be  used  on  a  laboratory  bench.  Inspect  the  instrument  and 
power cords for damage before first use. 

 
2.1 Electrical safety and preparation for use. 
 
Voltages  capable  of  causing  injury  or  death  are  present  in  this  instrument.    Use 
extreme caution whenever the instrument cover is removed.

 

 

Line Voltage 

This instrument is designed to operate with a 100 to 240 VAC, 47 to 63 Hz power source. 
This instrument is also capable of operating with a DC power source that can 12 
to +36 VDC at 2 Amperes when the DC option is acquired. 
 

Fuse 

A 1.0 Ampere 250V 5X10mm slow-blow fuse is used for 100 to 240 VAC operation. 
A 2.0 Ampere 250V 5X10mm slow-blow fuse is used for DC power operation. 
Only replace fuses with the same type and specifications.  
 

AC Power 

The instrument has a detachable three wire power cord for connection to a grounded AC 
power  source.  The  enclosure  of  the  unit  is  directly  connected  to  the  outlet  ground  to 
protect against electrical shock. Always use an outlet with a protective ground and do not 
disable  this  safety  mechanism.  Detaching  the  AC  power  cord  is  the  only  option  of 
disconnecting the unit from the AC mains supply. Make sure you have access to the rear 
panel or provide an external accessible AC disconnect means for your PPS-2RM-B. 
 

DC Power 

The PPS-2RM-B may be acquired with a DC power option to allow the instrument  to be 
powered  by  +12  to  +36  VDC,  2 Amp  source.  The  DC  supply  operation may  be used  as 
backup power to prevent loss of signal during power outages or as a mains supply source.

 

The switching from AC to DC supply operation is affected by a Schottky diode network and 
charge storage capacitors to ensure glitch free operation.  
 

If the PPS-2RM-B was acquired with the DC operation option 

a 6 pin DC connector will be 

available on the back panel.

 

 
 
 
 

Summary of Contents for PPS-2RM-B

Page 1: ...PPS 2RM B 1 PPS GENERATOR OPERATING MANUAL SPECTRADYNAMICS INC 1849 Cherry St Unit 2 Louisville CO 80027 Phone 303 665 1852 Fax 303 604 6088 www spectradynamics com SPECTRADYNAMICS INC ...

Page 2: ...PPS 2RM B 1 PPS Generator Operating Manual Copyright 2021 SpectraDynamics Inc All rights reserved PPS 2RM B R00 2021 MD ...

Page 3: ...ents 1 0 Introduction 1 2 0 Safety and preparation for use 2 2 1 Electrical 2 2 2 Instrument 4 3 0 Front panel description 5 4 0 Back panel description 6 5 0 Operation 7 6 0 Specifications 8 7 0 Warranty 9 ...

Page 4: ......

Page 5: ... 75 ohm cables The channel to channel delay differences are less than 500 ps The pulse per second generators are available in a 1U full rack mount enclosure that operate on 100 to 240 VAC Optional DC operation option is available for all models upon request PPS Options Part Number Description AC Operation DC Operation OPTION PPS 2RM A One generator two outputs Rear panel connectors PPS 2RM A DC On...

Page 6: ...operation Only replace fuses with the same type and specifications AC Power The instrument has a detachable three wire power cord for connection to a grounded AC power source The enclosure of the unit is directly connected to the outlet ground to protect against electrical shock Always use an outlet with a protective ground and do not disable this safety mechanism Detaching the AC power cord is th...

Page 7: ...e instrument and voids all warranties Please note that the power return pin 4 is NOT connected to the instrument case ground internally however both ground connections pin 4 and pin 6 are available at the DC power connector and may be connected together at this point The following specifications should be used to ensure optimum performance DC Supply voltage 12 to 36 VDC 2 Amps Line regulation 0 05...

Page 8: ...tings The PPS 2RM B pulse generator can be configured for different pulse widths The default configuration for the unit is 51 2 us pulse width To change the pulse width selection the unit must be unplugged from the power source Remove ONLY the top cover of the unit Locate the DIP switch labeled S5 on the module s to be configured The pulse width of the generator is a multiple of the clock period a...

Page 9: ...1 LED The LED will flash on the rising edge of the 1 PPS external synchronization signal from the first module EXT PPS2 LED The LED will flash on the rising edge of the 1 PPS external synchronization signal from the second module PPS1 LED The LED will flash on the rising edge of the 1 PPS output signal from the firts PPS generator module PPS2 LED The LED will flash on the rising edge of the 1 PPS ...

Page 10: ...d to an external 1 PPS signal Warning After a synchronization event this switch should be returned to the off or down position to prevent accidental synchronization Trigger When synchronization is enabled and a 1 PPS signal is present on the Ext 1 PPS input the 1 PPS synchronization event can be triggered by the push button Only one synchronization will occur for each trigger Phase Selects the ref...

Page 11: ...n the front panel will flash on the rising edge of each pulse To synchronize the output pulses to an external event connect the external reference pulse signal to the input labeled Ext 1 PPS on the back panel Enable the synchronization sequence by moving the Enable switch to the Enable position The Trigger button will arm the synchronization sequence to occur on the next rising edge at the Ext 1 P...

Page 12: ...MIN TYP MAX UNITS Rise time 10 90 1 5 2 ns Fall time 10 90 1 5 2 ns Differential delay Channel Channel 200 500 ps Impedance input output 50 10 Ohms Input High Level Input Low Level Input signal into 50 ohm load Input signal into 50 ohm load 2 0 7 5 0 8 V Output High Level Output Low Level 50 ohm load 50 ohm load 3 4 0 1 5 0 2 V Input Signal Level 1 MHz 5 MHz and 10 MHz 0 7 13 dBm External Sync Err...

Page 13: ...ion or any liability for direct indirect or consequential damages or delay Any improper use operation beyond capacity substitution of parts not approved by SDI or any alteration or repair by others in such manner as in SDI s reasonable judgement affects the product materially and adversely shall void this warranty No employee or representative of SDI is authorized to change this warranty in any wa...

Page 14: ......

Page 15: ...n voltage limits 2014 30 EU The Electromagnetic Compatibility Directive and its amending directives 2011 65 EU of the European Parliament and of the Council of 8 June 2011 on the restriction of the use of certain hazardous substances in electrical and electronic equipment Standards EN 55011 2009 A1 2010 Product family standard for Industrial Scientific and Medical ISM equipment EN 61010 1 2010 Saf...

Reviews: