34
CHAPTER 8.
Screen Messages and Troubleshooting
1. Warning Messages
Indication
Warning Description
[Warning: Low Battery]
At this time, battery is getting low but you can still
perform about 10 tests.
[Solution]
Replace the battery soon.
[Warning: Replace Battery]
Battery power is low.
If you press the Enter button after discharging the
battery, the battery icon will flash and then after ten
seconds the analyzer will turn off automatically.
[Solution]
Replace the battery immediately.
2. Error Messages
In certain circumstances error messages may appear on your display. Generally,
you should first try the solutions suggested for the respective error. If the problem
persists, please contact your local customer support and service center.
Indication
Warning Description
[Test Panel Error]
Defective test panel or the test panel is damaged or
inserted improperly.
[Solution]
Discard this test panel and test again using a new
test panel.
[Sample Error]
An insufficient amount of sample volume was
applied to the test panel.
[Solution]
Discard this test panel and test again using a new
test panel. When testing again, apply the proper
volume of sample mixture.