12
CHAPTER 2.
Testing the MultiCare Analyzer
1. Operating the Analyzer
Before using the analyzer for the first time, perform the following steps:
STEP 1.
Insert four batteries or connect the AC adapter jack.
STEP 1-1.
Insert four batteries
1)
Ensure the analyzer is switched off and turn it over.
2) Open the battery compartment lid by slightly pressing the tab towards
the center of the analyzer.
3)
Lift the lid upwards to remove it from the analyzer.
4) Insert four batteries into the compartment. Please note the orientation
of the “+” (battery head) and “–” terminals (flat end). Use only alkaline
batteries (1.5 V, AA).
NOTE
• If you think the display screen does not appear for long
enough, you can hold down the Enter button (the next time
you switch on the analyzer). The display screen is shown
during the time while the button is pressed.
•
After inserting or replacing batteries, confirm that the time
and date are set correctly. If they are not, reset the analyzer
before testing.
NOTE
Always replace all four batteries at the same time, because
the different capacities of batteries cause the analyzer to
malfunction. Do not use rechargeable batteries.
5) Close the battery compartment lid.
6) Turn the analyzer on to test the functionality of the new batteries.
7)
Check that the display is functioning correctly to prevent misinterpretations
due to defective display elements.
STEP 1-2.
Connect AC Adapter Jack
1)
AC adapter is not provided with MultiCare. Power supply adapter
5V/2A must be used.
2)
Connect the AC adapter jack to the AC jack port in the back of the analyzer.