background image

TB-9023

 Page 3 of 16

© 2017 DESCO INDUSTRIES, INC.

Employee Owned

Power Supply and Charger

Use only the power supply that comes with the EM Eye 

Meter. Do not use any other power supply as it may 

damage the meter.

Battery Care - Charging for the first time

To improve the battery life we recommend that you 

make three fully charge-discharge cycles. For example: 

fully charge the meter for two hours or overnight, 

then use the meter until it is fully discharged (do not 

recharge it at half-charge-life). Repeat this step two 

times. It would normally take two hours to fully charge 

the battery, but since the meter power is on while 

charging; it would extend the charge time to four hours. 

Succeeding charging would be at any duration. Use only 

the supplied charger.

Power On

Using your finger, press and hold any spot on the 

screen for approximately three seconds – do not use 

any tools to perform this action. If a new sensor was 

installed, a sequence of initialization will take place 

for the first time for approximately ten seconds. After 

that, the switch-on action will take approximately three 

seconds. The main display screen appears after the 

initialization and a beep will sound. The EM Eye Meter 

will then perform a battery check. If the battery is too 

low to provide reliable operation, it will not turn on. The 

screen will go white momentarily and then the meter will 

power up. After the initial power up, the meter will go 

directly to the main screen.

Connecting the Sensor Head

Features and Components

Figure 4. Connecting the sensors to the EM Eye Meter

ESD Sensor 

Assembly

EMF Sensor 

Assembly

RF Sensor 

Assembly

Figure 5. EM Eye features and components

Local directional antenna

Sensor module

SMA connectors

Alignment pin

Communication connectors
Alignment hole

Touchscreen display

MicroSD memory slot
Headphone jack

Reset pinhole

Power jack

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for CTC021

Page 1: ...e magnitude of most ESD events when in one of its filter modes electromagnetic fields EMF power density effective isotropic ally radiated power and radio frequency RF Made in the United States of Amer...

Page 2: ...microSD card while the power is on Turn off the meter before removing the microSD card Do not remove the sensor heads while the power is on Turn off the meter before removing and replacing the heads...

Page 3: ...e seconds do not use any tools to perform this action If a new sensor was installed a sequence of initialization will take place for the first time for approximately ten seconds After that the switch...

Page 4: ...n with no storage card in place PREPARING CARD FOR RECORDING DO NOT REMOVE PLEASE WAIT A new card is inserted or when a card is empty WARNING This card contains the data Remove the card or click here...

Page 5: ...t can occur A moving ungrounded cart may accumulate charges in its path As it approaches a metallic worktable and bumps into it a discharge may occur and may adversely affect nearby products or instru...

Page 6: ...damage and other problems For those who install and maintain antennae of radio transmitters and mobile phones the EM Eye Meter can provide readings of the field strength including peak and average va...

Page 7: ...C Employee Owned RF Signal Sensor Display and Controls EM Field Sensor Display and Controls Figure 11 Electromagnetic Field Sensor displays and controls Figure 12 RF Signal Sensor displays and control...

Page 8: ...R C button for approximately one second Figure 14 Using the Hold button Figure 15 Setting the time Battery The BATTERY symbol displays the approximate level of battery life remaining for use Full char...

Page 9: ...appears Press the up down arrows to scroll through functions Press the left or right arrows to set values and options Press the center button to return to the main screen Serial Number The serial numb...

Page 10: ...ure 26 Viewing stored data Figure 22 Navigating through different ESD Filter Models ESD Threshold The THRESHOLD in a form of an ARC LED bar displays the detection level of the ESD events detected When...

Page 11: ...ay For other memory card types please contact SCS Customer Service for verification Electromagnetic Field Sensor Power Density Meter and Effective Isotropically Radiated Power EIRP Meter Display and C...

Page 12: ...evel If any of the reference modes are enabled you should see the REF label shown on the main screen The bottom line of the display will now show the reference value of the peak signal Note The top li...

Page 13: ...od the hardware setup is shown At this point brightness off time and attenuation can be adjusted Attenuation ATT If the input to the antenna is big enough set the attenuation from 0 to 20dB When the a...

Page 14: ...llow you to download the data from the EM Eye Meter and save it to your PC 2 Install the EM Eye File Converter software 3 Remove the microSD card from the EM Eye Meter and insert it into your PC using...

Page 15: ...5 mV 15mV 1500mV Detection Resolution 1mV Threshold Display Resolution 1 10 Volts 10 990 Volts Raw input 1mV 1 15mV and 15mV 15 1500mV Distance Detection Range 1 3 cm 0 5 in 38 1 cm 15 0 in ESD Counte...

Page 16: ...with federal state and local requirements Regulatory Information FCC This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rule...