TB-9023
Page 15 of 16
© 2017 DESCO INDUSTRIES, INC.
Employee Owned
Calibration
The EM Eye Meter comes from the factory calibrated. It is recommended that you calibrate the EM Eye Meter once
a year at the factory.
Specifications
General
Audio Indicators
Speaker: Beep or analog audio with selectable volume
External Data Storage
microSD Card
Recording Interval
Peaks: 1 – 360 seconds
Average signals: 0.1 – 360 seconds
Display
Touch LCD
Size
65 W x 32 D x 105 L mm
Headphone Jack
3.5 mm (1/8 in.)
Power Supply/Charger
Input: 100-240V ~ 50-60 Hz 0.2A
Output: 5.0V 1000mA
Center: Positive, ID 1.3 mm; OD 3.5 mm
ESD Event Sensor
ESD Event Detection
Characterization Modes
Raw input for magnitude and estimated CDM, HBM, MM using proprietary algorithm
Raw Input Resolution
1mV (1mV-15mV) ; 15 mV (15mV-1500mV)
(Detection Resolution: 1mV)
Threshold (Display)
Resolution
1-10 Volts, 10-990 Volts
Raw input: 1mV (1-15mV) and 15mV (15-1500mV)
Distance Detection Range 1.3 cm (0.5 in.) – 38.1 cm (15.0 in.)
ESD Counter Range
0 – 32,767 counts (screen count)
Hardware Setup
Adjustable Brightness, Auto-off-time
Electromagnetic Field Sensor
Function Modes
Electromagnetic Field
Power Density Meter
EIRP Meter
Peak range
Min: 0.001 V/m (60 dBµV/m)
Max: 20.00 V/m (146.02 dBµV/m)
Min: 0.027 nW/cm
2
Max: 106.1 uW/cm
2
Min: 0.001W
Max: 5333 W
Average range
Min: 0.001 V/m
Max: 20.00 V/m
Min: 2.65 nW/m
2
Max: 1.06 W/m
2
Antenna Factor range
-40.0 to 40.0 dBm
Hardware setup
Brightness, Off-time, Antenna factor (AF)
Antenna options
Local, or remote (CTC111)
Record interval range
1 to 360 seconds
Record average range
0.1 to 360 seconds
Calibrated frequency
1 GHz
RF Signal Sensor
Unit modes
dBµV, dBm, mV
Dynamic range
Min: -55 dBm
Max: 0 dBm
ATT (Attenuation)
00 dB / 20 dB
www.
.com
1.800.561.8187