Version 3
|
7
2
Installation
Mini
WAVE
should be installed on the lab table that provides easy access considering the height of the lab’s operators. Or, a customized
Mini
WAVE
cart to fit up to 2 Modules & 1 Controller (010-600-014) is available.
The exhaust vent of the
Mini
WAVE
must be connected to a laboratory
exhaust system with a Positive Pull
40 cfm (1.1 cmm) or more to prevent
noxious fumes from entering the lab.
Installation Location
To install
Mini
WAVE
select a location that:
1. Is in close proximity to lab exhaust (a proper laboratory fume hood).
2. Provides a minimum clearance of 6 in. (15 cm) space in the back and 2 inches from each side.
3. Provides close access to a properly grounded electrical outlet.
4.
Mini
WAVE
should be operated on a stabilized constant voltage AC power supply, and the voltage must be within
+/-5% of the specified level.
5. Compressed air supply must be dry and oil free (5 Atms or 80 psi).
It is important to ensure that the exhaust hose is connected to laboratory exhaust.
Precautions to Avoid Possible Exposure to Excessive Microwave Energy
a. Do not attempt to operate this instrument with the lid open since open-lid operation.
can result in harmful exposure to microwave energy. It is important not to defeat or tamper with the safety interlocks.
b. Do not operate the instrument if it is damaged.
c. The instrument should not be adjusted or repaired by anyone except properly qualified service personnel.
d. Do not operate the microwave digestion system without a rack with samples in the cavity.
e. You should never defeat or tamper with the safety interlocks.
f. If the
Mini
WAVE
is used in a manner not specified by the manufacturer, the protection provided may be impaired.
NOTE