Version 3
|
29
Running display messages
DURING OPERATION THE FOLLOWING MESSAGES CAN BE SEEN:
Message
Description
1
Idle state
In the system idle state, the unit number is steady on the indicator.
2
Digestion initialization
The digestion starts by heating the magnetron filament and waiting for
stabilization. The unit number will flash while in this phase.
3
Ramping stage
During the temperature ramping stage, an animation is visible on the indicator: a
rolling pattern made of the sequence of segments A-B-C-D-E-F and repeated.
4
Holding stage
The hold phase has another animation pattern: alternating between two symbols:
a ‘u’ and a ‘n’ meaning that hold temperature over/under shoot is meant to be
under tight control. The segments sequence is ABF-CDE and repeats.
5
Digestion fail
If any critical anomaly shows during digestion, the digestion has to be halted. In
addition to the controller display emergency message, the message ‘F’ ‘a’ í’ ‘l’ is
sent on the indicator.
6
Explosion
In the event of an explosion, a flashing ‘H’ (for Halt) signals the user that it
occurred and the digestion is stopped.
7
Digestion done
When the digestion succeeds, the message ‘d’ ‘o’ ‘n’ ‘e’ appears on the
indicator.
8
Cooling phase
A big flashing ‘C’ indicates the cooling operation in progress.