12. With the assistance of eight additional people, for X500R systems, or nine additional people, for X500B systems,
move the mass spectrometer to the new location.
shows the position of the 10 people required to lift the X500B system. The person designated as
10
is at the front of the system, and must move aside to allow the system to be taken off and put onto the
bench. For X500R systems, only positions 1 through 9 are required.
Figure 7-14 Distribution of Operators (X500B System)
13. With the assistance of one additional person, move the roughing pump to the new location.
14. Install the left and right skirts.
15. Connect the vacuum hose, gas tubing, source exhaust tubing, power cable, ethernet cable, and TDC cable to
the mass spectrometer.
System User Guide
SCIEX X500 QTOF System
RUO-IDV-05-2334-E
218 / 234
Service and Maintenance Information