Sample Type Filter
Table 6-16 Sample Type Filter Descriptions
Description
Filter Type
Shows only Unknown samples, which are normal samples of unknown concentration. When
Standard samples are used, their concentration is back-calculated from the calibration curve
and reported in the Results Table as the Calculated Concentration.
Unknowns
Shows only samples of known concentration. These samples are used for the creation of the
calibration curve.
Standards
Shows only Quality Control samples. These samples of known concentration are used to
check the accuracy of the calibration curve but do not influence its actual construction.
Quality Controls
Shows only Blank samples. These are generally samples that contain the internal standard
compounds, if used, but no analytes, and that have been through the normal sample
preparation procedure. These samples are not used in the construction of the calibration
curve. To include them, select the Standard sample type and then set the Actual Concentration
at 0.
Blanks
Shows only Double Blank samples. These are samples with neither internal standards nor
analytes.
Double Blanks
Shows only Solvent samples. These are double blanks that have not been through the normal
sample work-up procedure.
Solvents
Acceptance Criteria
Use the acceptance criteria to define a qualifying row. A qualifying row matches the selected criteria for the
following filters:
• Pass: Show the rows that match the criteria that was defined in the processing method.
• Fail: Shows the rows that do not match the criteria that was defined in the processing method.
• Any: Shows all of the rows.
Table 6-17 Acceptance Criteria
Description
Label
Shows the user-defined criteria for integration quality using the value of the quality column
in the Results Table, asymmetry factor column, total width column, and the retention time
error columns (either %error column or absolute delta column).
Integration
acceptance
Shows the accuracy of the standard and quality control samples tolerance.
Accuracy
SCIEX X500 QTOF System
System User Guide
153 / 234
RUO-IDV-05-2334-E
Operating Instructions—Software