6
SM6016
2.
Technical Specifications
Below are the general specification and the accuracy specification of SM6016.
General Specifications
Function
Measurement
Parameters
Primary: L/C/R/Z/DCR
Secondary: D/Q/
θ
/ESR
Equivalent Mode
Series, Parallel
Auto LCR Function
Manual, Auto
Ranging Mode
Auto
Test Terminals
3-terminal,5-terminal
LCRZ
4 meas/sec,1.5meas/sec
Measurem
ent Speed
DCR
3 meas/sec,2.5meas/sec
Correction
Short, open
Tolerance Mode
1%
,
5%
,
10%
,
20%
Input Protection Fuse
0.1A / 63V
Interface
Mini-USB (virtual serial port)
Test Signal
Signal Frequency(0.02%
accuracy)
100Hz,*120Hz,1kHz,10kHz,100kHz
Test Signal Level(10%
accuracy)
0.3 Vrms, 0.6Vrms, 1Vrms
DCR signal
:
1Vdc
Output Impedance
100
Ω
Display
Display
LCD primary-secondary dual display
Backlight
Battery supply: when backlight is on,
luminance is reduced by half 15s later and
automatically turned off 30s later.
Powered by adapter: backlight off
until manually turned off
Count
Max. Counts of Primary Parameter:
40,000; D / Q /
θ
Min. resolution of
secondary parameter: 0.0001.
Basic Accuracy
0.1%
(
see performance check accuracy
specifications for details
)
Primary Range and
Resolution
See Performance check accuracy
Specifications