SM6016
23
Power-up State (PuP)
The “PuP” menu option allows user to configure the power-up state of the LCR
meter, allowing user to restore settings saved into internal EEPROM memory at
power-up. The storable settings are as follows:
Primary function mode (e.g. L/C/R)
Secondary function mode (e.g. D/Q)
Auto LCR
Series/parallel equivalent mode
Test frequency
Tolerance mode
Reference value for Tolerance mode
Measurement rate
In the utility menu, Press the
▲
and
▼
arrow keys to select “PrE” or “SEt”. PrE
means to preserve the previous setting while SEt means to save the current
parameter, that is, cover the original data.
NOTE:
Exiting mode decides whether to implement
SEt. At “Exiting without saving”,
SEt is ineffective; while at “Saving and Exiting”, SEt setting will be effective.
Default Setting: PrE
Configure and Save Power-up State
Procedure to configure and save power-up state is as follows:
1. Before entering into the utility menu, configure all the measurement
parameters firstly, such as frequency, primary and secondary parameters. If
the meter is currently in the utility menu, exit first and enter into utility menu
after measuring setup.
2. Press UTIL button for a long time to enter into utility menu.
3. Push the UTIL button to traverse through the utility menu until you see “PuP ”
on the primary display.
4. In order to save the current meter settings for power-up state into internal
memory, use either
▲
and
▼
button to change the settings so that the
secondary display shows “SEt”.
5. Press the UTIL button to check whether all desirable setups have already be
set. With all settings done, exit the menu by long press of the UTIL button.
6. The meter has now saved all current settings into internal memory. At next
power-up, the meter will turn on and recall the saved settings.
NOTE:
The meter allows one set of settings to be
stored into memory. Therefore, the
same procedure is used to overwrite previously stored settings into memory.
Prevent Overwrite of Stored Settings
In the utility menu, the “PuP” option default setting is always “ PrE ” . If it is required
to overwrite previously stored settings for power-up state, the option should change
to “SEt” . Therefore, when entering the utility menu, be sure not to change to “SEt” to
prevent overwriting any previously stored power-up settings.
Reset Default Settings (dEF)
The “dEF” option is used to reset the current measuring setup and optional settings