28
SM6016
7.
Quick Start Guide
7.1 CAUTION
Do not measure a capacitor that is not fully discharged. Connecting a
charged or partially charged capacitor to the input terminals will damage
the instrument.
When measuring an on- board component within a circuit, the circuit must
be powered off before connecting the test leads.
When used in a dusty environment, the instrument should be wiped and
cleaned regularly. The electrical conductivity carried by the accumulated
dust will eventually have an effect on the use of the meter.
Do not leave the instrument exposed to explosive, direct heated and
overheating environments.
Before removing the cover, ensure that none DUT is connected to the
meter and the instrument is disconnected from any circuit and is powered
OFF.
NOTE:
To achieve optimum precision, please see
“Clear Function” for details.
7.2 Inductance Measurement
1. Press the POWER button for a long time to turn on the instrument.
2. Press the PRI button until “L” is displayed on the screen to select
inductance measurement
3. Insert an inductor into test slots or connect a tested inductor through a
proper test accessory (e.g., banana plug-crocodile clip test leads, test
fixture or SMD test tweezers). See figures 13 below. Figure 5 shows the
optional 4-terminal test accessories.
4. Press the FREQ button until the desired test frequency is displayed on
screen.
5. Press the LEV button until the desired test level is displayed on screen.
6. Press the SEC button to select the desired secondary parameter.
7. Read the readings on LCD for inductance measured values.
Figure 13- Inductance Measurement