Instructions for NTEP models
Instructions for NTEP models
Instructions for NTEP models
Instructions for NTEP models
14
STORE START_OPTION N
↵
N = 0 Requires the ON/OFF button to be pressed to turn on
N = 1 Starts up as soon as power is applied. Balance does self test and then goes into normal weighing.
N = 2 Starts up as soon as power is applied. Bypasses self test and goes idirectly into normal weighing.
Note the single space between STORE and START_OPTION and a single space before N.
SETCLOCK MM/DD/YY HH:MM:SS
↵
Sets the real time clock to the date and time entered. The RTC is battery powered and will keep time for
months without external power.
Note the single space between SETCLOCK and MM/DD/YY and HH:MM:SS.
ASINT S
↵
S = Number of seconds between output
Note the single space between ASINT and S.
This is the Auto Send interval in seconds.
Check weighing over the remote operation.
Check weighing is also available from the I/O using the SETCHKWGT <high limit> <low limit> command.
Example:
SETCHKWGT 137.556 135.2
↵
places the balance in check weighing mode with the high limit set to 137.556 and the low limit set to 135.2
Note there is a space between SETCHKWGT and 137.556 and another space between 137.556 and 135.2
SETCHKWGT uses the current unit of measure. If the balance is in grams the <high limit> <low limit> num-
bers are expressed in grams as well.
Optional RS-232 Printers:
Please contact Scientech or one of our representatives for more information. A comprehensive manual is in-
cluded with each printer when purchased. All printers are set for a 9600 baud rate and even parity. To change
these settings please see the printer manual and the RS-232 Front Panel Configuration section. The balance
display contents can be sent to the printer at any time by pressing the balance front panel SEND button.
Many other accessories are available. Please contact Scientech at 303-444-1361 or 800-525-0522 or email us at
[email protected] for more information.