background image

Instructions for NTEP models

Instructions for NTEP models

Instructions for NTEP models

Instructions for NTEP models    

11 

RS-232 Interface: 

Note: 

Scientech  balances  are  DTE  (computer)  devices.  If  communicating  to  another  DTE  device  (ie 
computer), a Scientech RS-232 cable, PN11897, or normal null modem cable is required. If you 
are communicating to a DCE device (printer), a serial cable is required. Scientech offers a serial 
cable, PN11367,  for your convenience. 

 

Your balance is equipped with a bi-directional RS-232 compatible inter-
face. The RS-232 connector is a 9 pin D subminiature connector located 
on  the  balance  rear  panel.  The  pin  out  and  pin  descriptions  are  shown 
here.  

RS-232 Specifications: 

Type 

 

 

EIA-RS-232C, DTE 

Method   

 

Half-duplex, Asynchronous Transmission, Bi-directional 

Format   

 

4800, 9600 and 19,200 baud rate selectable (default is 9600) 

 

 

 

Code:  ASCII 

 

 

 

For parity Even or Odd:   Data bit 7, Stop bit 1. 

 

 

 

For parity none:   

Data bit 8, Stop bit 1. 

 

RS-232 Front Panel Configuration:

 

Before  using  the  RS-232  feature  of  your  balance,  it  is  necessary  to  configure  the  interface  properly  for  your 
application.  The  RS-232  Configuration  Mode  allows  you  to  set  baud  rate,  auto  send,  continuous  send,  send 
when stable and display blanking. The balance is configured with defaults at the factory. The default modes is 
9600 baud, no parity. 

 

It is not necessary to follow the entire configuration procedure if only one parameter is to be changed. Use step 
1 to enter the menu and press MODE/UNITS to scroll through the different parameters. When the desired  
parameter is displayed the SEND key will scroll through the settings for  that  parameter.  Pressing the  
RANGE /FUNCTION at any time will save the settings and return to normal weighing mode. 

 

Your terminal program and the balance must have the same baud rate and parity to work. If you choose NONE 
for parity please set your terminal program to 8 data bits, 7 data bits for EVEN or ODD parity. 
 
 
 
To enter the RS-232 configuration mode via the front panel, proceed as follows: 

  

User Action

 

Balance Response

 

Press ON/OFF and SEND simultaneously. 

The  current  baud  rate  br4800,  br9600,  or 
br19,200 is displayed. 

Press the SEND button until the desired baud rate 
is displayed.   

 

Press the MODE/UNITS button. 

The current parity even, odd or none is displayed. 

Press the SEND button until the desired parity is 
displayed.   

 

Press the MODE/UNITS button. 

Auto send AS ON or AS OFF is displayed 

Press the SEND button until the desired auto send 
mode is displayed.   

 

Press the MODE/UNITS button. 

Continuous send CS ON or CS OFF is displayed 

Summary of Contents for NTEP

Page 1: ...Instructions for NTEP models Instructions for NTEP models Instructions for NTEP models Instructions for NTEP models SCIENTECH NTEP CERTIFIED 11777 D Scientech inc com ...

Page 2: ...ation CAL 1 Using an External Weight and meteroological sealing 4 Good Measurement Practices 5 Equipment 5 Environment 5 Operator 5 Level the Balance 6 Connect to an Electrical Outlet 6 Turn the balance on 6 Front Panel Controls 7 The ON OFF I O Button 7 The RANGE FUNCTION Button 7 The ZERO Buttons 7 The MODE UNITS Button 7 The SEND Button 7 Balance Functions and Features 8 Front Panel Parts Count...

Page 3: ...ent The environment in which your balance is used is very important Air movement temperature changes vibrations direct sunlight etc influences the performance of high precision balances Therefore place your balance on a solid sturdy surface that is free of air currents vibration and not in direct sunlight The surface should not be magnetic and should be located away from doors windows heaters air ...

Page 4: ...uld be connected to an electrical outlet for at least one 1 hour prior to calibration and leveled according to the Unpacking and Setup section The balance was shipped with the calibration seal un locked When the scale calibration seal is un locked the display will show CAL in the upper left corner of the LCD The scale is not legal for trade when CAL is displayed If CAL is visible you can skip step...

Page 5: ...ts are in thermal equilibrium with the balance producing a better measurement Environment 1 The more stable your environment the better your measurement results Changes in temperature pressure and humidity affect balance performance and weight stability Ideal room conditions are 20º C with a relative humidity between 45 and 60 Fluctuations in temperature should not exceed 1º C per hour Humidity fl...

Page 6: ...into the balance s rear panel receptacle Then plug the power supply into an AC outlet Note Scientech recommends that the balance be plugged into an electrical outlet at all times This ensures the balance is always warmed up and ready for use You can leave the balance on as it consumes 5 Watts Thumbwheel turns but does not move up or down A shaft within the thumbwheel is extended or re tracted This...

Page 7: ...eros When a weighing that has been zeroed out is removed from the weighing pan a negative reading is displayed To return the display to zeros press one of the ZERO buttons Note The balance will not zero during a disturbance The MODE UNITS Button Pressing the MODE UNITS button starts the unit of weight cycle as follows For NP models grams g and milligrams mg only For NJ models grams g carats ct pen...

Page 8: ...ividual piece weight is 30mg If the initial 10 piece sample size has a cumulative weight less than 300mg when the ZERO button is pressed the display will blank and the LO annunciator will flash for three seconds and the part counting mode will terminate and the balance will return to the normal weighing mode Percent Weighing ZSP400 NP only User Action Balance Response 1 Place an empty container on...

Page 9: ...The balance then returns to normal operation Capacity Tracker The capacity tracker provides a graphic display of the used and unused portions of the weighing range and is shown in Figure 7 Each segment represents 10 of the balance s total capacity As 10 of the balance s capacity is used the first segment will illu minate As 20 of the balance s capacity is reached the second segment will light and ...

Page 10: ...d place an empty graduated flask on the balance weighing pan tare the balance then pour the liquid into the graduated flask Divide the displayed weight in grams by the volume of the liquid The result is the mass per unit volume of the liquid in g cc Security Bracket A security bracket is provided on the balance s rear panel as a convenient method of securing the balance User Action Balance Respons...

Page 11: ... send when stable and display blanking The balance is configured with defaults at the factory The default modes is 9600 baud no parity It is not necessary to follow the entire configuration procedure if only one parameter is to be changed Use step 1 to enter the menu and press MODE UNITS to scroll through the different parameters When the desired parameter is displayed the SEND key will scroll thr...

Page 12: ...l The balance expects ASCII character instruction strings and responds with ASCII character data strings All characters with an ASCII value less than the space character decimal 32 of hex 20 are ignored by the balanc es USB The RS 232 and USB ports share the same internal communication port All commands described by the RS 232 are valid for the USB as well The balance utilizes a CP 2102 serial bri...

Page 13: ...at the interval specified by ASINT Turns auto send on or off When auto send is on a time interval must be entered See SETA SINT ASINT n Auto Send Interval n Seconds Sets the auto send interval in Seconds CSON or CSOFF Output the current weight at every 200mS Turns continuous send on or off SSON or SSOFF Output the current weight when the OK annunciator ap pear in the display Turns send when stable...

Page 14: ...ion Check weighing is also available from the I O using the SETCHKWGT high limit low limit command Example SETCHKWGT 137 556 135 2 places the balance in check weighing mode with the high limit set to 137 556 and the low limit set to 135 2 Note there is a space between SETCHKWGT and 137 556 and another space between 137 556 and 135 2 SETCHKWGT uses the current unit of measure If the balance is in g...

Page 15: ...ble Drafts air currents are present Vibrations are present Install windscreen relocate bal ance Isolate balance from vibration relocate balance Weight readings are incorrect Balance is out of calibration Balance is not level Display was not zeroed before weight was placed on pan Object being weighed is touching windscreen or balance case top Unit of weight selection is incor rect Recalibrate balan...

Page 16: ...2 C to 30 C Relative Humidity 80 for temperatures up to 31ºC decreasing linearly to 50 at 40ºC Power Supply Automatic switching 115 230 Volts 50 60 Hz 10 Windscreen Analytical Windscreen Calibration Weight grams 150 100 400 200 Housing 7 5 W x 11 25 D x 3 25 H Weight pounds 9 5 Optional Specifications for ZSP150 NJ Capacity d e ZSP400 NJ Capacity d e ZSP400 NP Capacity d e Grams 150g d 001g e 01g ...

Page 17: ...that box in a larger box with at least four inches of packing material between the two boxes Scientech does not assume responsibility for under packed items Please include the name and phone number of the person to contact regarding repair question s Normally products are repaired and shipped within 5 working days after their arrival at the product service facility This is an average time and coul...

Page 18: ...Instructions for NTEP models Instructions for NTEP models Instructions for NTEP models Instructions for NTEP models 18 ...

Reviews: