USER MANUAL
(Instruction for Use)
SW 2.9
EMC Guidance and Manufacturer’s Declaration
Version 2.9.7 dated 2020-12-23
Page 69 of 75
_
NOTE
:
“The
EMISSIONS
characteristics of this equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11
class B is normally required) this equipment might not offer adequate protection to radio-
frequency communication services. The user might need to take mitigation measures, such as
relocating or re-orienting the equipment.”
N
OTE
:
S
ERVICE RECOMMENDATION
:
To prevent affect to the Emissions and Immunity of the equipment throughout the expected
service life it is recommended to use the product always with its complete and original enclosure,
with original parts from the manufacturer and to use the product according to the EMC
instructions in the IFU.
Table 1: ELECTROMAGNETIC EMMISSIONS
Guidelines and declaration of conformity – electromagnetic emissions
The Carriazo-Pendular is intended to be used in the electromagnetic environment specified
below. The customer or the user of the Carriazo-Pendular should assure that it is used in such
an environment.
Electromagnetic
emission
Compatibility
Electromagnetic environment - standards
RF emission
CISPR 11
Group 1
The Carriazo-Pendular uses RF energy for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emission
CISPR 11
Conducted
emissions
(Interference
voltage):
Class B
Radiated
emissions:
Class A
NOTE:
The EMISSIONS characteristics of this
equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in
a residential environment (for which CISPR 11 class
B is normally required) this equipment might not
offer adequate protection to radio-frequency
communication services. The user might need to
take mitigation measures, such as relocating or re-
orienting the equipment.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions IEC
61000-3-3
Complies