USER MANUAL
(Instruction for Use)
SW 2.9
Device / System Description
Version 2.9.7 dated 2020-12-23
Page 38 of 75
_
(1) Serial Connector:
(2) Foot pedal socket:
Safe screwed connection to
foot pedals
(3) Secondary fuse F3
(4) Accu charging fuse F4
(5) Primary fuses F1, F2
and
Power Supply Socket
with
indicator of the selected input voltage.
The standard setting of 220-240V can be changed to
110-120V, (indication: 115V or 230V), (refer to chapter
).
4.3.8
Foot Pedal
(1) Forward Pedal:
Forward motion of cutting head with blade oscillation. Motion stops when the pedal is
released.
(2)
Reverse Pedal
:
Reverse motion of the cutting head without oscillation of the blade. Motion stops when
the pedal is released.
(3)
VAC Pedal:
Press and release the pedal once to activate the vacuum; press and release the pedal a
second time to stop the vacuum. The vacuum can also be activated directly with the VAC-
button on the console.
(Version starting in October 2012)
1
2
5
4
3