USER MANUAL
(Instruction for Use)
SW 2.9
Technical Data
Version 2.9.7 dated 2020-12-23
Page 66 of 75
_
10
TECHNICAL DATA
10.1
Electronic Console
Classification:
Medical device class:
IIa
(Directive 93/42/EEC Annex IX)
Electrical protection:
Class I
(IEC 60601-1)
Housing protection class
:
Cosole safety standards:
IP 20
(IEC 60529)
Foot switch safety standards:
IP 67
(IEC 60529)
Overall device data
Dimensions:
28 x 23 x 21 cm (length x width x height)
Weight:
8.8 kg
Electric specifications
Power supply:
100-120 VAC 50/60Hz or 220-240 VAC 50/60 Hz
Power consumption
0.6 A @120V AC, 0.3 A @ 230 V AC
Fuses F1, F2:
120V: 2 x T3.15AL / 250 V
240V: 2 x T1.6AL / 250 V
Fuse F3:
T 3.15 AL / 250 V
Fuse F4
T 3.15 AL / 250 V
Environmental Conditions
Temperature:
15 - 35 °C
(only Carriazo-Pendular, consider “Room and
Installation Requirements for laser surgery”)
Humidity:
up to 60 %
(only Carriazo-Pendular, consider “Room and
Installation Requirements for laser surgery”)
Atmosphere:
860 hPa - 1060 hPa
Displayed vacuum:
+/- 5%
10.2
Suction Rings
Suction rings made of surgical stainless steel.
Suction rings are available for different flap diameters.
10.3
Cutting Head
Cutting heads made of surgical stainless steel.
Cutting heads are available for different flap thicknesses.