-5-
-6-
3.2.Display
Display of unit
Display of measuring value.
Display of peak hold mode
Display the measuring value with sign and 4 digits numbers. Compression force: plus, Tension force: minus.
(It’ s available to switch plus/minus with the setting of function (f01)).
Reverse display is available.
3.2.2.Numeric Display
Display the units. In case of overload, “OVR” is shown. “PWR” appears to notify that there is 1 minute before power off.
3.2.3.Unit Display
Display of unit
Overload
3.2.4.Peak hold mode display
In case of the voltage of
internal nickel hydride
battery decrease, “ LO
BAT “ has turned on and
off. Please connect AC
adapter to charge the
battery.
D u r i n g t h e c h a r g e
of battery, “BAT” is
s h o w n . E v e n i f t u r n
the power off during
the charge “BAT” has
displayed.
“PEAK” has turned on
when plus peak hold
mode,
※
“PEAK” has turned on
when minus peak hold
mode.
※
Newton
Depend on the condition of the force gauge, the following display is shown.
Kilogram
Gram
1 minut before
power off.
3.2.1.Part names
Reverse display is available.
※ Pleaes discern plus peak hold mode and minus peak hold mode with or without “– “.
Kg/g, lb/oz are decided by rating capacity.
Pound
Ounce
-5-
-6-
3. Part names and functions
3.1.Main Unit
①
Sensor shaft
Force bolt (Push/Pull Force)
②
Display
Display the load and the unit.
③ PEAK key
・ Switch standard measuring mode, + peak hold mode, or – peak hold mode.
・ Tracking ON/OFF.
・ It is used for function setting.
④ UNIT key
・ Switch the unit (N ⇒ Kg ⇒ lb(oz) ⇒ N)
・ It is used for reverse the display.
・ Tracking ON/OFF.
・ It is used for function setting.
⑤ ZERO key
・ Tare at standard measuring mode (not display a peak value).
・ It is used for function setting.
・ Press the ZERO key under the peak hold mode (display the peak value), clear the peak value. In this case,
tare does not perform. If you need tare, press the peak key to change into standard measuring mode,
then press the Zero key.
⑥ POWER key
・ Press the POWER and release to turn ON/OFF.
・ It is used for reverse the display.
・ Tracking ON/OFF.
・ It is used for function setting.
⑦
AC adapter port
Supply the electricity through the AC adapter.
⑧
Hanger attachement bolt
Attache the hanger with this bolt.
⑨
Threaded hole
Use this threaded hole in order to attach the FGJN with a stand.
② Display
① Sensor shaft
③ PEAK key
④UNIT key
⑤ZERO key
⑥POWER key
Model number
⑨ Threaded hole
⑧ Hanger attachment bolt
⑦ AC adapter port