E n d o P i l o t
*related to the additional module “CoPilot”
- 30 -
609 2201 v12 www.schlumbohm.de
Part I – Continued …
Maintenance:
C.
Material: disposable tissues, Quick-acting-disinfection-spray FD 333
1
www.duerrdental.com
, spray-adapter for ISO-contra-angles and oil spray F1
(product of W&H www.wh.com)
-Process: spray the oil continuously through the contra angle for about 1 sec, remove
surplus oil with a disposable tissue. Turn the contra angle downwards (the head),
attach the motor (no oil may flow into the motor). Keep the motor running for about 30
sec to remove surplus oil. Start with minimum rotational peed and increase to
maximum speed. If contaminants leak out, repeat the whole cleansing, disinfection,
drying, and maintenance process
Alternatively we recommend oil-service with W&H Assistina
(please see the manufacturer’s instruction)
Finally wipe motor and contra angle (separately) thoroughly.
Inspection and checking of the
function:
All parts: Inspect the parts for signs of damage and wear.
Do not use damaged components.
Packing:
Contra angle: separately in standardized sterile-bag for steam sterilization
Size of packing has to be sufficient.
Caution! Motor: The motor must not be sterilised
Sterilisation:
Steam sterilization (fractioned vacuum) autoclave device class B: 5 min at 134°C, 2
bar. Treatment duration should be sufficient for the potential germs, if necessary,
choose the prion-program (for example: Melag Vakuklave 43 B)
Storage:
No special requirements.
Additional information:
When sterilising several instruments in one cycle, maximum load capacity of the
autoclave must not be exceeded.
Contact to manufacturer:
For questions please contact your dealer or the manufacturer www.schlumbohm.de
The instructions stated above have been validated as SUITABLE by the medical device manufacturer for the
preparation of a medical product prior to its re-usage. The person who does the treatment is responsible that the
treatment that actually took place with the equipment, material and personnel used in the treatment institution leads to
the desired results. To ensure this it is normally necessary to validate and routinely supervise the process. Each
deviation from the instructions provided should be evaluated carefully by the person executing the treatment with
regard to effects and negative consequences.
1
FD 333 , Quick-acting-disinfection-spray (
www.duerrdental.com
)