6.1.
Selection of the Ultrasonic output level*
The level of intensity is freely selectable on the touch display.
Please select an appropriate level of intensity. At the beginning, always select a low output level and then increase it up
to the level needed. Do not activate the instruments if they are still on air (without contact to the tooth). The ultrasonic
device is intended for an intermittent (interrupted) use. In order to avoid heating, do not run the device longer than 2
minutes at maximum power or 4 minutes at minimum power.
Ensure sufficient cooling-off times:
2 minutes use at maximum performance, 6 minutes cooling-off.
4 minutes use at minimum performance, 6 minutes cooling-off.
6.2.
Selection of Run-Time*
Rinsing liquids often require pre-set application times. The selection of the run-time enables to work with defined
activation times. The CoPilot run-time can be freely selected from 1 to 240 seconds. After the countdown it stops
automatically with an audible signal!
Always select a very low ultrasonic power level at first and increase the energy only when needed.
The ultrasonic device is intended for intermittent (interrupted) use, always ensure sufficient cooling-off times (see above).
6.3. Ultrasonic-Instrument-Selection
With the
Instrument
button (resp. name of the selected instrument), a choice of ultrasonic instruments can be called.
The relevant performance parameters are pre. The choice of the instrument is stored and retained, even if you turn off
the CoPilot.