Pilot²
- 26 -
4.5.
Setup motor
The settings that apply to the entire preparation are
chosen under this menu.
It is possible to individually change the selected file’s
parameters. Select the
File data
. The File setup opens
(see below).
You can configure further settings for the apex
measurement during motor operation under
Apex
setup
.
You can test the motor with the contra-angle handpiece
under
Calibrate
. Leave the motor and the contra-angle
handpiece in the holder and start the calibration.
4.5.1. File data
All US registered, pre-programmed nickel titanium files with a standard ISO shaft, tools-
coupling ISO 14457:2017 can be used. The device offers pre-programmed standard settings,
provided from the tool manufacturers.
The File setup menu offers a variety of setting options. The device allows all parameters such
as speed, torque, operating mode, etc. to be changed individually. Please contact the file
manufacturer first, if you want to change the Parameters. Parameters that differ from the
instrument manufacturer’s specifications may lead to file breakage and other damage.
Under the Movement menu item, you can select:
•
Twist (right rotation with ‘Shake loose’ when jamming)
•
Twist off (right movement, motor stops in the event of
overload)
•
Left rotation
•
Reciprocal function
Note:
Please keep in mind that the file manufacturers reserve
the right to make changes and customizations to the
instrument’s characteristic values. The data stored in this
device has been defined with great care in accordance
with manufacturer recommendations. The user can
individually customize future changes or import such
changes via an update.