Pilot²
- 22 -
Setup button (see chapter 7)
3. Manual apex length determination
4. Motor system for NiTi files
5. BackFill gun DownPack handpiece
6. Ultrasonic function
Battery status display
3. Manual apex length determination
In this menu you can pre-probe the canal manually, i.e.
with a file guided by the hand. Use the lip clip (4b) and the
file clamp (4e) for this.
The marker (horizontal line) determines the position in the
root canal where the ‘Auto-Stop’ function is reached
during mechanical preparation. The manufacturer has
already set the marker. The user (if desired) can, however,
change the setting directly on the display by moving the
bar (by tapping on it with the finger). With this function it is
possible to transfer the X-ray verified position of a pilot
instrument to the display.
The marker’s setting remains unchanged until the device
is switched off. If the device is switched on again, the line
is reset to the default value.
Do not place the measuring cables on electrically
conductive surfaces, as external voltages could be
transferred to the device.
Use the button to access the settings in the Apex setup menu.