8
Technical Data
8.3
Defibrillation pulse
Page 84
A
rt.
no.
: 0-4
8
-00
13 Rev. m
FRED easy
Shock delivery
• With the orange button
(in semi-automatic or manual mode)
• Via disposable pads applied to the patient in an anterior-anterolateral or anteri-
or-posterior position
Safety discharge when:
• A non shockable rhythm has been detected
• The shock is not delivered within the 20 seconds after charging
• An electrode problem is detected
• Battery voltage is insufficient
• The device is defective
• The device is turned off.
Defibrillation pad connection
BF type
Defibrillation electrodes
Adult pads
Paediatric pad
Electrode cable, 2 m in length
• 80 cm² active surface
• 42 cm² active surface
VF/VT detection
Conditions for ECG analysis
Shock recommendation
VF/VT detection is based only on the ECG signal.
Minimum amplitude for signals suitable for analysis > 0.15 mV, signals < 0.15 mV
are considered asystole.
For VF and VT (VT > 150 B/min).
VF
VT
Non shockable
rhythms
NSR/N/Asystole
Shock
148
100
2
No Shock
2
2
1395
Performance criteria
Sensitivity > 90 %
Sensitivity > 75 %
Specificity > 95 %
Observed performance
98.67
%
98.04
%
99.86
%